Point shear wave ultrasound elastography with Esaote compared to real-time 2D shear wave elastography with supersonic imagine for the quantification of liver stiffness.
Mulazzani, L; Salvatore, V; Ravaioli, F; Allegretti, G; Matassoni, F; Granata, R; Ferrarini, A; Stefanescu, H; Piscaglia, Fabio.
Afiliação
  • Mulazzani L; Department of Medical and Surgical Sciences, University of Bologna, Bologna, Italy.
  • Salvatore V; Unit of Internal Medicine, Hospital S.Orsola-Malpighi, via Albertoni 15, 40138 Bologna, Italy.
  • Ravaioli F; Department of Medical and Surgical Sciences, University of Bologna, Bologna, Italy.
  • Allegretti G; Unit of Internal Medicine, Hospital S.Orsola-Malpighi, via Albertoni 15, 40138 Bologna, Italy.
  • Matassoni F; Department of Medical and Surgical Sciences, University of Bologna, Bologna, Italy.
  • Granata R; Unit of Internal Medicine, Hospital S.Orsola-Malpighi, via Albertoni 15, 40138 Bologna, Italy.
  • Ferrarini A; Department of Medical and Surgical Sciences, University of Bologna, Bologna, Italy.
  • Stefanescu H; Unit of Internal Medicine, Hospital S.Orsola-Malpighi, via Albertoni 15, 40138 Bologna, Italy.
  • Piscaglia F; Department of Medical and Surgical Sciences, University of Bologna, Bologna, Italy.
J Ultrasound ; 20(3): 213-225, 2017 Sep.
Article em En | MEDLINE | ID: mdl-28900522
ABSTRACT

PURPOSE:

Different shear wave elastography (SWE) machines able to quantify liver stiffness (LS) have been recently introduced by various companies. The aim of this study was to investigate the agreement between point SWE with Esaote MyLab Twice (pSWE.ESA) and 2D SWE with Aixplorer SuperSonic (2D SWE.SSI). Moreover, we assessed the correlation of these machines with Fibroscan in a subgroup of patients.

METHODS:

A total of 81 liver disease patients and 27 subjects without liver disease accessing the ultrasound lab were considered. Exclusion criteria were liver nodules, BMI >35, and severe comorbidities. LS was sampled from the same intercostal space with both pSWE.ESA and 2D SWE.SSI and values were tested with Lin's analysis and Bland-Altman analysis (B&A). Agreement between each SWE machine and Fibroscan was assessed in 26 liver disease patients with Spearman correlation.

RESULTS:

Precision and accuracy between pSWE.ESA and 2D SWE.SSI were, respectively, 0.839 and 0.999. B&A showed a mean of only -0.2 kPa, with no systematic deviation between the techniques and limits of agreement at -11.6 and 11.3 kPa. Spearman's rho correlation versus Fibroscan was 0.849 for pSWE.ESA and 0.878 for 2D SWE.SSI. The relationship became less strict in the higher range of LS (≥15.2 kPa), corresponding to cirrhosis.

CONCLUSION:

The overall degree of concordance of pSWE.ESA and 2D SWE.SSI in measuring LS resulted remarkable, also when compared with Fibroscan. The less strict correlation for patients with LS in the higher range would not affect the staging of disease as such patients are anyhow classified as cirrhotic.
Assuntos
Palavras-chave

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Técnicas de Imagem por Elasticidade / Fígado / Hepatopatias Tipo de estudo: Diagnostic_studies / Observational_studies / Prevalence_studies / Risk_factors_studies Limite: Adult / Aged / Aged80 / Female / Humans / Male / Middle aged Idioma: En Revista: J Ultrasound Ano de publicação: 2017 Tipo de documento: Article País de afiliação: Itália

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Técnicas de Imagem por Elasticidade / Fígado / Hepatopatias Tipo de estudo: Diagnostic_studies / Observational_studies / Prevalence_studies / Risk_factors_studies Limite: Adult / Aged / Aged80 / Female / Humans / Male / Middle aged Idioma: En Revista: J Ultrasound Ano de publicação: 2017 Tipo de documento: Article País de afiliação: Itália