A retrospective study of radiation dose measurements comparing different cath lab X-ray systems in a sample population of patients undergoing percutaneous coronary intervention for chronic total occlusions.
McNeice, Andrew H; Brooks, Matthew; Hanratty, Colm G; Stevenson, Michael; Spratt, James C; Walsh, Simon J.
Afiliação
  • McNeice AH; Department of Cardiology, Belfast Health and Social Care Trust, c/o Level 9, Belfast City Hospital, Lisburn Road, Belfast, BT9 7AB, Northern, Ireland.
  • Brooks M; Department of Cardiology, Edinburgh Royal Infirmary, Edinburgh, Scotland.
  • Hanratty CG; Department of Cardiology, Belfast Health and Social Care Trust, c/o Level 9, Belfast City Hospital, Lisburn Road, Belfast, BT9 7AB, Northern, Ireland.
  • Stevenson M; Department of Epidemiology & Public Health, Queen's University, University Road, Belfast, Northern Ireland.
  • Spratt JC; Department of Cardiology, Edinburgh Royal Infirmary, Edinburgh, Scotland.
  • Walsh SJ; Forth Valley Hospital, Larbert, Scotland.
Catheter Cardiovasc Interv ; 92(3): E254-E261, 2018 09 01.
Article em En | MEDLINE | ID: mdl-29411518
ABSTRACT

OBJECTIVES:

A retrospective study was performed to investigate if the generation of X-ray system used was an independent factor for radiation dose in chronic total occlusion (CTO) percutaneous coronary intervention (PCI).

BACKGROUND:

PCI procedures for CTOs are known to be associated with higher doses of radiation. The authors suspected progressive reductions in radiation doses for CTO PCI as newer X-ray systems were introduced into clinical practice.

METHODS:

Procedures performed over a five-year period by three interventional cardiologists were retrospectively reviewed. Five different X-ray systems were used across three hospital sites. These included Axiom Artis and Coroskop HIP (both Siemens), Innova (GE), Allura Xper FD 10, and Allura Clarity FD 10 (both Philips). Procedural and demographic data including body mass index (BMI; kg/m2 ), fluoroscopy time (min), and dose area product (DAP; cGycm2 ) were collated for each procedure. Statistical analysis was performed to compare the influence each X-ray system would have on DAP values after BMI and fluoroscopy time were controlled for.

RESULTS:

In total, 860 procedures were analyzed. Mean fluoroscopy time was 40.00 ± 19.99 min, mean BMI was 29.90 ± 5.13 kg/m2 , mean DAP 11,980 ± 7,947 cGycm2 . Log values of DAP were used to normalize results in a general linear model. A significant statistical difference in DAP between X-ray systems was demonstrated after fluoroscopy time and BMI were controlled for (P ≤ 0.001).

CONCLUSION:

There is a significant impact on DAP values resulting from the generation of X-ray system used, measured during PCI for CTOs, with the most modern systems producing the lowest radiation doses.
Assuntos
Palavras-chave

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Doses de Radiação / Cateterismo Cardíaco / Radiografia Intervencionista / Angiografia Coronária / Exposição à Radiação / Vasos Coronários / Oclusão Coronária / Intervenção Coronária Percutânea Tipo de estudo: Etiology_studies / Observational_studies / Prognostic_studies / Risk_factors_studies Limite: Humans País/Região como assunto: Europa Idioma: En Revista: Catheter Cardiovasc Interv Assunto da revista: CARDIOLOGIA Ano de publicação: 2018 Tipo de documento: Article País de afiliação: Irlanda

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Doses de Radiação / Cateterismo Cardíaco / Radiografia Intervencionista / Angiografia Coronária / Exposição à Radiação / Vasos Coronários / Oclusão Coronária / Intervenção Coronária Percutânea Tipo de estudo: Etiology_studies / Observational_studies / Prognostic_studies / Risk_factors_studies Limite: Humans País/Região como assunto: Europa Idioma: En Revista: Catheter Cardiovasc Interv Assunto da revista: CARDIOLOGIA Ano de publicação: 2018 Tipo de documento: Article País de afiliação: Irlanda