Lower airway obstruction due to a massive clot resulting from late bleeding following mini-tracheostomy tube insertion and subsequent clot removal and re-intubation.
Inoue, Hiroshi; Ito, Jun; Uchida, Hiroaki; Morita, Mariko; Masuda, Takahiko; Yamaya, Kazuhiro; Hata, Masaki; Kato, Shigeaki.
Afiliação
  • Inoue H; 1Department of Anesthesia and Critical Care Medicine, Sendai Kousei Hospital, 4-15 Hirosemachi, Aoba-ku, Sendai City, Miyagi 9800873 Japan.
  • Ito J; 1Department of Anesthesia and Critical Care Medicine, Sendai Kousei Hospital, 4-15 Hirosemachi, Aoba-ku, Sendai City, Miyagi 9800873 Japan.
  • Uchida H; 1Department of Anesthesia and Critical Care Medicine, Sendai Kousei Hospital, 4-15 Hirosemachi, Aoba-ku, Sendai City, Miyagi 9800873 Japan.
  • Morita M; Department of Anesthesia, Minamisoma Municipal General Hospital, 2-54-6 Haramachi-ku Takamimachi, Minami-Souma City, Fukushima 9750033 Japan.
  • Masuda T; 3Department of Cardiovascular Surgery, Sendai Kousei Hospital, 4-15 Hirosemachi, Aoba-ku, Sendai City, Miyagi 9800873 Japan.
  • Yamaya K; 3Department of Cardiovascular Surgery, Sendai Kousei Hospital, 4-15 Hirosemachi, Aoba-ku, Sendai City, Miyagi 9800873 Japan.
  • Hata M; 3Department of Cardiovascular Surgery, Sendai Kousei Hospital, 4-15 Hirosemachi, Aoba-ku, Sendai City, Miyagi 9800873 Japan.
  • Kato S; 4Cardiovascular Center, Sendai Kousei Hospital, 4-15 Hirosemachi, Aoba-ku, Sendai City, Miyagi 9800873 Japan.
JA Clin Rep ; 3(1): 16, 2017.
Article em En | MEDLINE | ID: mdl-29457060
ABSTRACT

BACKGROUND:

Easier to perform than the conventional procedure, mini-tracheostomy (MT) is widely used in the operating room or intensive care unit to remove sputum or other obstructions of the upper airway. This option, however, does carry the risk of various complications, including malposition, disposition, bleeding, and subcutaneous emphysema. Here, we report a case of endotracheal tube obstruction due to a massive clot resulting from late bleeding around the insertion site of an MT tube. This necessitated removal of the endotracheal tube together with the clot followed tube re-introduction. CASE PRESENTATION The patient was an 85-year-old man in whom an MT tube had been inserted 6 days earlier following aortic replacement surgery. On re-admittance to our intensive care unit, large amounts of hemosputum and clotting were observed around the insertion site of the tube. The MT tube was subsequently removed and tracheal intubation performed. Ventilation via the endotracheal tube proved impossible, however, and cardiac arrest ensued. Fiberoptic bronchoscopy revealed that the endotracheal tube was completely obstructed by a massive clot. Therefore, we immediately pushed the clot toward the right main bronchus to secure ventilation via the left lung. After many attempts to remove the massive clot, including suction and grasping with basket forceps, it was successfully dislodged by replacing the endotracheal tube with a new one while maintaining oxygenation by one-lung ventilation. Any small fragments of the clot that still remained were then removed by suction under fiberoptic bronchoscopy.

CONCLUSIONS:

Here, we report a case of endotracheal tube obstruction due to a clot derived from very late (6 days) bleeding after insertion of an MT tube. The patient was successfully rescued by replacing the clot-bearing endotracheal tube with a new one. This experience suggests that the intensive care physician should be aware of the potential risk of clot retention in endotracheal tubes after the elapse of several days.