Microdosimetry at the CATANA 62 MeV proton beam with a sealed miniaturized TEPC.
Conte, V; Bianchi, A; Selva, A; Petringa, G; Cirrone, G A P; Parisi, A; Vanhavere, F; Colautti, P.
Afiliação
  • Conte V; INFN Laboratori Nazionali di Legnaro, viale dell'Università 2, 35020 Legnaro, Italy. Electronic address: conte@lnl.infn.it.
  • Bianchi A; INFN Laboratori Nazionali di Legnaro, viale dell'Università 2, 35020 Legnaro, Italy; Belgian Nuclear Research Centre, SCK·CEN, Boeretang 200, 2400 Mol, Belgium; UHasselt, Faculty of Engineering Technology, Centre for Environmental Sciences, Nuclear Technology Center, Agoralaan, 3590 Diepenbeek, Belg
  • Selva A; INFN Laboratori Nazionali di Legnaro, viale dell'Università 2, 35020 Legnaro, Italy.
  • Petringa G; INFN Laboratori Nazionali del Sud, via S. Sofia 62, 95123 Catania, Italy.
  • Cirrone GAP; INFN Laboratori Nazionali del Sud, via S. Sofia 62, 95123 Catania, Italy.
  • Parisi A; Belgian Nuclear Research Centre, SCK·CEN, Boeretang 200, 2400 Mol, Belgium.
  • Vanhavere F; Belgian Nuclear Research Centre, SCK·CEN, Boeretang 200, 2400 Mol, Belgium.
  • Colautti P; INFN Laboratori Nazionali di Legnaro, viale dell'Università 2, 35020 Legnaro, Italy.
Phys Med ; 64: 114-122, 2019 Aug.
Article em En | MEDLINE | ID: mdl-31515010
ABSTRACT
A new mini-TEPC with cylindrical sensitive volume of 0.9 mm in diameter and height, and with external diameter of 2.7 mm, has been developed to work without gas flow. With such a mini counter we have measured the physical quality of the 62 MeV therapeutic proton beam of CATANA (Catania, Italy). Measurements were performed at six precise positions along the Spread-Out Bragg Peak (SOBP) 1.4, 19.4, 24.6, 29.0, 29.7 and 30.8 mm, corresponding to positions of clinical relevance (entrance, proximal, central, and distal-edge of the SOBP) or of high lineal energy transfer (LET) increment (distal-dose drop off). Without refilling the microdosimeter with new gas, the measurements were repeated at the same positions 4 months later, in order to study the stability of the response in sealed-mode operation. From the microdosimetric spectra the frequency-mean lineal energy y-F and the dose-mean lineal energy y-D were derived and compared with average LET values calculated by means of Geant4 simulations. The comparison points out, in particular, a good agreement between microdosimetric y-D and the total dose-average LET¯d, which is the average LET of the mixed radiation field, including the contribution by nuclear reactions.
Assuntos
Palavras-chave

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Prótons / Radiometria / Microtecnologia Tipo de estudo: Health_economic_evaluation Idioma: En Revista: Phys Med Assunto da revista: BIOFISICA / BIOLOGIA / MEDICINA Ano de publicação: 2019 Tipo de documento: Article

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Prótons / Radiometria / Microtecnologia Tipo de estudo: Health_economic_evaluation Idioma: En Revista: Phys Med Assunto da revista: BIOFISICA / BIOLOGIA / MEDICINA Ano de publicação: 2019 Tipo de documento: Article