Your browser doesn't support javascript.
loading
Comparison of Automated CT Perfusion Softwares in Evaluation of Acute Ischemic Stroke.
Xiong, Yunyun; Huang, Chiwen C; Fisher, Marc; Hackney, David B; Bhadelia, Rafeeque A; Selim, Magdy H.
Afiliação
  • Xiong Y; Department of Neurology, Stroke Division, Beth Israel Deaconess Medical Center, Harvard Medical School, Boston, Massachusetts. Electronic address: yxiong@bidmc.harvard.edu.
  • Huang CC; Department of Radiology, Beth Israel Deaconess Medical Center, Harvard Medical School, Boston, Massachusetts; Department of Radiology, Wan Fang Hospital, Taipei Medical University, Taipei, Taiwan; Department of Radiology, School of Medicine, College of Medicine, Taipei Medical University, Taipei, Ta
  • Fisher M; Department of Neurology, Stroke Division, Beth Israel Deaconess Medical Center, Harvard Medical School, Boston, Massachusetts.
  • Hackney DB; Department of Radiology, Beth Israel Deaconess Medical Center, Harvard Medical School, Boston, Massachusetts.
  • Bhadelia RA; Department of Radiology, Beth Israel Deaconess Medical Center, Harvard Medical School, Boston, Massachusetts.
  • Selim MH; Department of Neurology, Stroke Division, Beth Israel Deaconess Medical Center, Harvard Medical School, Boston, Massachusetts.
J Stroke Cerebrovasc Dis ; 28(12): 104392, 2019 Dec.
Article em En | MEDLINE | ID: mdl-31562038
ABSTRACT
BACKGROUND AND

PURPOSE:

Automated imaging software is integral to decision-making in acute ischemic stroke (AIS) during extended time windows. RAPID software is the most widely used and has been validated in landmark endovascular trials. Olea software is another commercially available and FDA-approved software, but has not been studied in AIS trials. We aimed to compare the diagnostic utility and accuracy of RAPID and Olea in everyday clinical practice outside of a clinical trial.

METHODS:

We analyzed prospectively-collected data from a consecutive cohort of 141 patients with suspected AIS who underwent computed tomography perfusion upon presentation followed by diffusion-weighted magnetic resonance imaging (DWI-MRI) within 24-48 hours. Core infarct was defined as the region with a relative cerebral blood flow (rCBF) less than 30% on RAPID and rCBF less than 40% on Olea (default settings). We also evaluated rCBF less than 30% on Olea to match RAPID's default setting. Infarct volume on DWI-MRI was measured using a semiautomated segmentation method.

RESULTS:

Twenty-one patients were excluded; 14 due to poor bolus tracking and/or motion artifact, and 7 due to software failure. The software failure rate was 4.7% [6/127] with RAPID versus .78% [1/127] with Olea (P = .12). For the remaining 120 patients, the sensitivity and specificity for detecting an acute infarct were 40.5% and 97.6% for RAPID; 50.6% and 85.4% for Olea; and for detecting large infarcts (≥70 mL on DWI-MRI) 73.7% and 81.2% for RAPID; 73.7% and 68.3% for Olea. Core infarct volume on RAPID was more closely correlated with DWI-MRI infarct volume (rho = .64) than Olea (rho = .42).

CONCLUSIONS:

Our head-to-head comparison of these 2 commonly-used softwares in the clinical setting elucidates the pros and cons of their use to guide decision-making for AIS management in the acute setting.
Assuntos
Palavras-chave

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Software / Interpretação de Imagem Radiográfica Assistida por Computador / Isquemia Encefálica / Circulação Cerebrovascular / Acidente Vascular Cerebral / Imagem de Perfusão / Tomografia Computadorizada Multidetectores Tipo de estudo: Observational_studies / Prognostic_studies / Risk_factors_studies Limite: Aged / Aged80 / Female / Humans / Male / Middle aged Idioma: En Revista: J Stroke Cerebrovasc Dis Assunto da revista: ANGIOLOGIA / CEREBRO Ano de publicação: 2019 Tipo de documento: Article

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Software / Interpretação de Imagem Radiográfica Assistida por Computador / Isquemia Encefálica / Circulação Cerebrovascular / Acidente Vascular Cerebral / Imagem de Perfusão / Tomografia Computadorizada Multidetectores Tipo de estudo: Observational_studies / Prognostic_studies / Risk_factors_studies Limite: Aged / Aged80 / Female / Humans / Male / Middle aged Idioma: En Revista: J Stroke Cerebrovasc Dis Assunto da revista: ANGIOLOGIA / CEREBRO Ano de publicação: 2019 Tipo de documento: Article