Clinical Use of a New High-Sensitivity Cardiac Troponin I Assay in Patients with Suspected Myocardial Infarction.
Clin Chem
; 65(11): 1426-1436, 2019 11.
Article
em En
| MEDLINE
| ID: mdl-31570633
ABSTRACT
BACKGROUND:
We aimed to validate the clinical performance of the high-sensitivity cardiac troponin I [VITROS® Immunodiagnostic Products hs Troponin I (hs-cTnI-VITROS)] assay.METHODS:
We enrolled patients presenting to the emergency department with symptoms suggestive of acute myocardial infarction (AMI). Final diagnoses were centrally adjudicated by 2 independent cardiologists considering all clinical information, including cardiac imaging first, using serial hs-cTnT-Elecsys (primary analysis) and, second, using hs-cTnI-Architect (secondary analysis) measurements in addition to the clinically used (hs)-cTn. hs-cTnI-VITROS was measured at presentation and at 1 h in a blinded fashion. The primary objective was direct comparison of diagnostic accuracy as quantified by the area under the ROC curve (AUC) of hs-cTnI-VITROS vs hs-cTnT-Elecsys and hs-cTnI-Architect, and in a subgroup also hs-cTnI-Centaur and hs-cTnI-Access. Secondary objectives included the derivation and validation of an hs-cTnI-VITROS-0/1-h algorithm.RESULTS:
AMI was the adjudicated final diagnosis in 158 of 1231 (13%) patients. At presentation, the AUC for hs-cTnI-VITROS was 0.95 (95% CI, 0.93-0.96); for hs-cTnT-Elecsys, 0.94 (95% CI, 0.92-0.95); and for hs-cTnI-Architect, 0.92 (95% CI, 0.90-0.94). AUCs for hs-cTnI-Centaur and hs-cTnI-Access were 0.95 (95% CI, 0.94-0.97). Applying the derived hs-cTnI-VITROS-0/1-h algorithm (derivation cohort n = 519) to the validation cohort (n = 520), 53% of patients were ruled out [sensitivity, 100% (95% CI, 94.1-100)] and 14% of patients were ruled in [specificity, 95.6% (95% CI, 93.4-97.2)]. Patients ruled out by the 0/1-h algorithm had a survival rate of 99.8% at 30 days. Findings were confirmed in the secondary analyses using the adjudication including serial measurements of hs-cTnI-Architect.CONCLUSIONS:
The hs-cTnI-VITROS assay has at least comparable diagnostic accuracy with the currently best validated hs-cTnT and hs-cTnI assays. CLINICALTRIALSGOV IDENTIFIER NCT00470587.
Texto completo:
1
Coleções:
01-internacional
Base de dados:
MEDLINE
Assunto principal:
Bioensaio
/
Troponina I
/
Infarto do Miocárdio
Tipo de estudo:
Clinical_trials
/
Diagnostic_studies
/
Observational_studies
/
Prognostic_studies
Limite:
Aged
/
Aged80
/
Female
/
Humans
/
Male
/
Middle aged
Idioma:
En
Revista:
Clin Chem
Assunto da revista:
QUIMICA CLINICA
Ano de publicação:
2019
Tipo de documento:
Article
País de afiliação:
Suíça