Your browser doesn't support javascript.
loading
Image Quality and Activity Optimization in Oncologic 18F-FDG PET Using the Digital Biograph Vision PET/CT System.
van Sluis, Joyce; Boellaard, Ronald; Dierckx, Rudi A J O; Stormezand, Gilles N; Glaudemans, Andor W J M; Noordzij, Walter.
Afiliação
  • van Sluis J; Department of Nuclear Medicine and Molecular Imaging, University of Groningen, University Medical Center Groningen, Groningen, The Netherlands j.van.sluis@umcg.nl.
  • Boellaard R; Department of Nuclear Medicine and Molecular Imaging, University of Groningen, University Medical Center Groningen, Groningen, The Netherlands.
  • Dierckx RAJO; Department of Nuclear Medicine and Molecular Imaging, University of Groningen, University Medical Center Groningen, Groningen, The Netherlands.
  • Stormezand GN; Department of Nuclear Medicine and Molecular Imaging, University of Groningen, University Medical Center Groningen, Groningen, The Netherlands.
  • Glaudemans AWJM; Department of Nuclear Medicine and Molecular Imaging, University of Groningen, University Medical Center Groningen, Groningen, The Netherlands.
  • Noordzij W; Department of Nuclear Medicine and Molecular Imaging, University of Groningen, University Medical Center Groningen, Groningen, The Netherlands.
J Nucl Med ; 61(5): 764-771, 2020 05.
Article em En | MEDLINE | ID: mdl-31628214
ABSTRACT
The first Biograph Vision PET/CT system (Siemens Healthineers) was installed at the University Medical Center Groningen. Improved performance of this system could allow for a reduction in activity administration or scan duration. This study evaluated the effects of reduced scan duration in oncologic 18F-FDG PET imaging on quantitative and subjective imaging parameters and its influence on clinical image interpretation.

Methods:

Patients referred for a clinical PET/CT scan were enrolled in this study, received a weight-based 18F-FDG injected activity, and underwent list-mode PET acquisition at 180 s per bed position (s/bp). Acquired PET data were reconstructed using the vendor-recommended clinical reconstruction protocol (hereafter referred to as "clinical"), using the clinical protocol with additional 2-mm gaussian filtering (hereafter referred to as "clinical+G2"), and-in conformance with European Association of Nuclear Medicine Research Ltd. (EARL) specifications-using different scan durations per bed position (180, 120, 60, 30, and 10 s). Reconstructed images were quantitatively assessed for comparison of SUVs and noise. In addition, clinically reconstructed images were qualitatively evaluated by 3 nuclear medicine physicians.

Results:

In total, 30 oncologic patients (22 men, 8 women; age 48-88 y [range], 67 ± 9.6 y [mean ± SD]) received a single weight-based (3 MBq/kg) 18F-FDG injected activity (weight 45-123 kg [range], 81 ± 15 kg [mean ± SD]; activity 135-380 MBq [range], 241 ± 47.3 MBq [mean ± SD]). Significant differences in lesion SUVmax were found between the 180-s/bp images and the 30- and 10-s/bp images reconstructed using the clinical protocols, whereas no differences were found in lesion SUVpeak EARL-compliant images did not show differences in lesion SUVmax or SUVpeak between scan durations. Quantitative parameters showed minimal deviation (∼5%) in the 60-s/bp images. Therefore, further subjective image quality assessment was conducted using the 60-s/bp images. Qualitative assessment revealed the influence of personal preference on physicians' willingness to adopt the 60-s/bp images in clinical practice. Although quantitative PET parameters differed minimally, an increase in noise was observed.

Conclusion:

With the Biograph Vision PET/CT system for oncologic 18F-FDG imaging, scan duration or activity administration could be reduced by a factor of 3 or more with the use of the clinical+G2 or the EARL-compliant reconstruction protocol.
Assuntos
Palavras-chave

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Fluordesoxiglucose F18 / Tomografia por Emissão de Pósitrons combinada à Tomografia Computadorizada / Neoplasias Tipo de estudo: Guideline / Qualitative_research Limite: Aged / Aged80 / Female / Humans / Male / Middle aged Idioma: En Revista: J Nucl Med Ano de publicação: 2020 Tipo de documento: Article País de afiliação: Holanda

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Fluordesoxiglucose F18 / Tomografia por Emissão de Pósitrons combinada à Tomografia Computadorizada / Neoplasias Tipo de estudo: Guideline / Qualitative_research Limite: Aged / Aged80 / Female / Humans / Male / Middle aged Idioma: En Revista: J Nucl Med Ano de publicação: 2020 Tipo de documento: Article País de afiliação: Holanda