Burst Pressures in Eyes with Clear Corneal Incisions Treated with ReSure Glue.
Shehata, Ahmed E M; Kambhampati, Siva P; Wang, Jiangxia; Soiberman, Uri S.
Afiliação
  • Shehata AEM; Cornea Division, Wilmer Eye Institute, Johns Hopkins University School of Medicine, Baltimore, MD, USA.
  • Kambhampati SP; Center for Nanomedicine, Wilmer Eye Institute, Johns Hopkins University School of Medicine, Baltimore, MD, USA.
  • Wang J; Department of Biostatistics, Johns Hopkins University Bloomberg School of Public Health, Baltimore, MD, USA.
  • Soiberman US; Cornea Division, Wilmer Eye Institute, Johns Hopkins University School of Medicine, Baltimore, MD, USA.
J Ophthalmol ; 2021: 6691489, 2021.
Article em En | MEDLINE | ID: mdl-33614168
ABSTRACT

PURPOSE:

This study aims to measure burst pressures in 3 mm clear corneal incisions sealed with ReSure, a biodegradable hydrogel sealant, and to compare it to traditional 10-0 nylon sutures and unsealed controls.

DESIGN:

An ex vivo animal study.

METHODS:

3 mm clear corneal incisions were performed in rabbit eyes (ex vivo). The burst pressure was determined, and then, the incisions were sealed with either ReSure glue or a single 10-0 nylon suture. Burst pressure measurements were repeated.

RESULTS:

Fourteen eyes were included. The median burst pressure in the suture-control group (7 eyes) prior to suture application was 7 mmHg (range 0-45); the median burst pressure in the 7 glue-controls was 36 mmHg (range 5-61, p = 0.08 for the comparison of the two control groups). The median burst pressure in the glue group was 93 mmHg (range 39-129, p = 0.043 when compared to glue-control). The median burst pressure in the suture group was 158 mmHg (range 70-180, p = 0.018 when compared to suture-control). There was no statistically significant difference in burst pressure values between the glue and suture groups (p = 0.08).

CONCLUSION:

In this study, ReSure glue applied to 3 mm clear corneal incisions provided sufficient resistance to elevated intraocular pressure when compared to controls. The results of this study suggest that ReSure glue may be comparable to a single 10-0 nylon suture in resisting fluid egress during the early postoperative period.