Your browser doesn't support javascript.
loading
[Agreement analysis and result calibration by ordinary least product regression between bioelectrical impedance analysis and dual energy X-ray absorptiometry].
Song, Xiaoyun; Wang, Huijun; Su, Chang; Ouyang, Yifei; Lyu, Zhimei; Wei, Nan; Li, Yan; Ma, Yuxia; Li, Qingxia; Zhang, Bing.
Afiliação
  • Song X; National Institute for Nutrition and Health, Chinese Center for Disease Control and Prevention, Beijing 100050, China.
  • Wang H; National Institute for Nutrition and Health, Chinese Center for Disease Control and Prevention, Beijing 100050, China.
  • Su C; National Institute for Nutrition and Health, Chinese Center for Disease Control and Prevention, Beijing 100050, China.
  • Ouyang Y; National Institute for Nutrition and Health, Chinese Center for Disease Control and Prevention, Beijing 100050, China.
  • Lyu Z; National Institute for Nutrition and Health, Chinese Center for Disease Control and Prevention, Beijing 100050, China.
  • Wei N; National Institute for Nutrition and Health, Chinese Center for Disease Control and Prevention, Beijing 100050, China.
  • Li Y; National Institute for Nutrition and Health, Chinese Center for Disease Control and Prevention, Beijing 100050, China.
  • Ma Y; School of Public Health, Hebei Medical University, Shijiazhuang 050017, China.
  • Li Q; School of Public Health, Hebei Medical University, Shijiazhuang 050017, China.
  • Zhang B; National Institute for Nutrition and Health, Chinese Center for Disease Control and Prevention, Beijing 100050, China.
Wei Sheng Yan Jiu ; 50(3): 426-431, 2021 May.
Article em Zh | MEDLINE | ID: mdl-34074365
ABSTRACT

OBJECTIVE:

To analyze agreement of body fat percentage measured by bioelectrical impedance analysis(BIA) and dual energy X-ray absorptiometry(DXA) in Chinese adults, and to calibrate BIA measurement by ordinary least product regression(OLP).

METHODS:

Volunteers aged 18-65 were recruited to measure total body fat percentage and trunk fat percentage by BIA(Tanita BC-601 F) and DXA(Hologic Discovery QDR), respectively. Agreement between the two methods was compared by OLP. Bootstrap method was used to do the internal validation of the calibration equation. The calibration effect was verified by the agreement analysis of the calibrated BIA result and the DXA result.

RESULTS:

Before calibration, all 95%CIs of the intercepts of the OLP regression lines did not include 0, and all 95%CIs of the slopes of the OLP regression lines did not include 1. 00. After calibration, all 95% CIs of the intercepts of the OLP regression lines included 0, and all 95% CIs of the slopes included 1. 00.

CONCLUSION:

There are fixed bias and proportion bias between BIA and DXA in measuring total body fat percentage and trunk fat percentage in Chinese men and women. Calibration should be considered when body fat percentage is measured by BIA.
Assuntos
Palavras-chave

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Composição Corporal / Povo Asiático Tipo de estudo: Prognostic_studies Limite: Adult / Female / Humans / Male Idioma: Zh Revista: Wei Sheng Yan Jiu Assunto da revista: SAUDE PUBLICA Ano de publicação: 2021 Tipo de documento: Article País de afiliação: China

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Composição Corporal / Povo Asiático Tipo de estudo: Prognostic_studies Limite: Adult / Female / Humans / Male Idioma: Zh Revista: Wei Sheng Yan Jiu Assunto da revista: SAUDE PUBLICA Ano de publicação: 2021 Tipo de documento: Article País de afiliação: China