Robust analysis of space-, time-, and energy-resolved soft x-ray measurements of magnetically confined fusion plasmas (invited).
VanMeter, P D; Delgado-Aparicio, L F; Reusch, L M; Den Hartog, D J.
Afiliação
  • VanMeter PD; Department of Physics, University of Wisconsin-Madison, Madison, Wisconsin 53706, USA.
  • Delgado-Aparicio LF; Princeton Plasma Physics Laboratory (PPPL), Princeton, New Jersey 08540, USA.
  • Reusch LM; Department of Physics, University of Wisconsin-Madison, Madison, Wisconsin 53706, USA.
  • Den Hartog DJ; Department of Physics, University of Wisconsin-Madison, Madison, Wisconsin 53706, USA.
Rev Sci Instrum ; 92(5): 053549, 2021 May 01.
Article em En | MEDLINE | ID: mdl-34243280
ABSTRACT
A novel compact multi-energy soft x-ray (ME-SXR) diagnostic based on the PILATUS3 100K x-ray detector has been developed in collaboration between the Princeton Plasma Physics Laboratory and the University of Wisconsin-Madison and tested on the Madison Symmetric Torus (MST) reversed-field pinch. This solid-state photon-counting detector consists of a two-dimensional array of ∼100 000 pixels for which the lower photon absorption cutoff energy can be independently set, allowing it to be configured for a unique combination of simultaneous spatial, spectral, and temporal resolution of ∼1 cm, 100 eV, and 500 Hz, respectively. The diagnostic is highly versatile and can be readily adapted to diverse plasma operating conditions and scientific needs without any required downtime. New results from improved-confinement and quasi-single helicity plasmas in the MST demonstrate how the detector can be applied to study multiple aspects of the evolution of magnetically confined fusion-grade plasmas. These include observing the evolution of thermal emissivity, characterizing the energy of mid-Z excitation lines, extracting the Te profile, and observing the evolution of non-thermal populations. A technique for integrating the ME-SXR diagnostic into an integrated data analysis framework based on Bayesian inference is also presented. This allows ME-SXR measurements to be combined with data for complementary diagnostics in order to simultaneously infer Te and nZ from all available information.

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Idioma: En Revista: Rev Sci Instrum Ano de publicação: 2021 Tipo de documento: Article País de afiliação: Estados Unidos

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Idioma: En Revista: Rev Sci Instrum Ano de publicação: 2021 Tipo de documento: Article País de afiliação: Estados Unidos