Reliability of mechanical ventilation during continuous chest compressions: a crossover study of transport ventilators in a human cadaver model of CPR.
Orlob, Simon; Wittig, Johannes; Hobisch, Christoph; Auinger, Daniel; Honnef, Gabriel; Fellinger, Tobias; Ristl, Robin; Schindler, Otmar; Metnitz, Philipp; Feigl, Georg; Prause, Gerhard.
Afiliação
  • Orlob S; Division of Anaesthesiology for Cardiovascular Surgery and Intensive Care Medicine, Department of Anaesthesiology and Intensive Care Medicine, Medical University of Graz, Auenbruggerplatz 29, 8036, Graz, Austria. simon.orlob@medunigraz.at.
  • Wittig J; Institute for Emergency Medicine, University Hospital Schleswig-Holstein, Campus Kiel, Arnold-Heller-Straße 3, 24105, Kiel, Germany. simon.orlob@medunigraz.at.
  • Hobisch C; Medical University of Graz, Auenbruggerplatz 2, 8036, Graz, Austria.
  • Auinger D; Division of General Anaesthesiology, Emergency- and Intensive Care Medicine, Department of Anaesthesiology and Intensive Care Medicine, Medical University of Graz, Auenbruggerplatz 29, 8036, Graz, Austria.
  • Honnef G; Division of General Anaesthesiology, Emergency- and Intensive Care Medicine, Department of Anaesthesiology and Intensive Care Medicine, Medical University of Graz, Auenbruggerplatz 29, 8036, Graz, Austria.
  • Fellinger T; Division of General Anaesthesiology, Emergency- and Intensive Care Medicine, Department of Anaesthesiology and Intensive Care Medicine, Medical University of Graz, Auenbruggerplatz 29, 8036, Graz, Austria.
  • Ristl R; Centre for Medical Statistics, Informatics and Intelligent Systems, Medical University of Vienna, Spitalgasse 23, 1090, Vienna, Austria.
  • Schindler O; Centre for Medical Statistics, Informatics and Intelligent Systems, Medical University of Vienna, Spitalgasse 23, 1090, Vienna, Austria.
  • Metnitz P; Department of Internal and Respiratory Medicine, Intensive Care Unit Enzenbach, State Hospital Graz II, Hörgas 30, 8112, Gratwein, Austria.
  • Feigl G; Division of General Anaesthesiology, Emergency- and Intensive Care Medicine, Department of Anaesthesiology and Intensive Care Medicine, Medical University of Graz, Auenbruggerplatz 29, 8036, Graz, Austria.
  • Prause G; Division of Macroscopic and Clinical Anatomy, Medical University of Graz, Harrachgasse 21, 8010, Graz, Austria.
Scand J Trauma Resusc Emerg Med ; 29(1): 102, 2021 Jul 28.
Article em En | MEDLINE | ID: mdl-34321068
ABSTRACT

BACKGROUND:

Previous studies have stated that hyperventilation often occurs in cardiopulmonary resuscitation (CPR) mainly due to excessive ventilation frequencies, especially when a manual valve bag is used. Transport ventilators may provide mandatory ventilation with predetermined tidal volumes and without the risk of hyperventilation. Nonetheless, interactions between chest compressions and ventilations are likely to occur. We investigated whether transport ventilators can provide adequate alveolar ventilation during continuous chest compression in adult CPR.

METHODS:

A three-period crossover study with three common transport ventilators in a cadaver model of CPR was carried out. The three ventilators 'MEDUMAT Standard²', 'Oxylog 3000 plus', and 'Monnal T60' represent three different interventions, providing volume-controlled continuous mandatory ventilation (VC-CMV) via an endotracheal tube with a tidal volume of 6 mL/kg predicted body weight. Proximal airflow was measured, and the net tidal volume was derived for each respiratory cycle. The deviation from the predetermined tidal volume was calculated and analysed. Several mixed linear models were calculated with the cadaver as a random factor and ventilator, height, sex, crossover period and incremental number of each ventilation within the period as covariates to evaluate differences between ventilators.

RESULTS:

Overall median deviation of net tidal volume from predetermined tidal volume was - 21.2 % (IQR 19.6, range [- 87.9 %; 25.8 %]) corresponding to a tidal volume of 4.75 mL/kg predicted body weight (IQR 1.2, range [0.7; 7.6]). In a mixed linear model, the ventilator model, the crossover period, and the cadaver's height were significant factors for decreased tidal volume. The estimated effects of tidal volume deviation for each ventilator were - 14.5 % [95 %-CI -22.5; -6.5] (p = 0.0004) for 'Monnal T60', - 30.6 % [95 %-CI -38.6; -22.6] (p < 0.0001) for 'Oxylog 3000 plus' and - 31.0 % [95 %-CI -38.9; -23.0] (p < 0.0001) for 'MEDUMAT Standard²'.

CONCLUSIONS:

All investigated transport ventilators were able to provide alveolar ventilation even though chest compressions considerably decreased tidal volumes. Our results support the concept of using ventilators to avoid excessive ventilatory rates in CPR. This experimental study suggests that healthcare professionals should carefully monitor actual tidal volumes to recognise the occurrence of hypoventilation during continuous chest compressions.
Assuntos
Palavras-chave

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Respiração Artificial / Reanimação Cardiopulmonar Tipo de estudo: Clinical_trials / Prognostic_studies Limite: Adult / Humans Idioma: En Revista: Scand J Trauma Resusc Emerg Med Assunto da revista: MEDICINA DE EMERGENCIA / TRAUMATOLOGIA Ano de publicação: 2021 Tipo de documento: Article País de afiliação: Áustria

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Respiração Artificial / Reanimação Cardiopulmonar Tipo de estudo: Clinical_trials / Prognostic_studies Limite: Adult / Humans Idioma: En Revista: Scand J Trauma Resusc Emerg Med Assunto da revista: MEDICINA DE EMERGENCIA / TRAUMATOLOGIA Ano de publicação: 2021 Tipo de documento: Article País de afiliação: Áustria