Your browser doesn't support javascript.
loading
Comparison of a ceiling-mounted 3D flat panel detector vs. conventional intraoperative 2D fluoroscopy in plate osteosynthesis of distal radius fractures with volar locking plate systems.
Seuthe, Raphael; Seekamp, Andreas; Kurz, Bodo; Pfarr, Julian; Schaefer, Jost Philipp; Peh, Simon; Lippross, Sebastian.
Afiliação
  • Seuthe R; Department of Trauma and Orthopedic Surgery, University Medical Center of Schleswig-Holstein, Campus Kiel, Arnold-Heller-Str. 3, 24105, Kiel, Germany.
  • Seekamp A; Department of Trauma and Orthopedic Surgery, University Medical Center of Schleswig-Holstein, Campus Kiel, Arnold-Heller-Str. 3, 24105, Kiel, Germany.
  • Kurz B; Department of Anatomy, Christian-Albrechts-University, Kiel, Germany.
  • Pfarr J; Department of Radiology and Neuroradiology, University Medical Center of Schleswig-Holstein, Campus Kiel, Kiel, Germany.
  • Schaefer JP; Department of Radiology and Neuroradiology, University Medical Center of Schleswig-Holstein, Campus Kiel, Kiel, Germany.
  • Peh S; Department of Trauma and Orthopedic Surgery, University Medical Center of Schleswig-Holstein, Campus Kiel, Arnold-Heller-Str. 3, 24105, Kiel, Germany.
  • Lippross S; Department of Trauma and Orthopedic Surgery, University Medical Center of Schleswig-Holstein, Campus Kiel, Arnold-Heller-Str. 3, 24105, Kiel, Germany. sebastian.lippross@uksh.de.
BMC Musculoskelet Disord ; 22(1): 924, 2021 Nov 02.
Article em En | MEDLINE | ID: mdl-34727909
ABSTRACT

OBJECTIVES:

To compare intraoperative 3D fluoroscopy with a ceiling-mounted flat panel detector in plate osteosynthesis of distal radius fractures (AO/OTA 2R3C1.2) with volar locking plate systems to conventional 2D fluoroscopy for detection of insufficient fracture reduction, plate misplacement and protruding screws.

METHODS:

Using a common volar approach on 12 cadaver forearms, total intraarticular distal radius fractures were induced, manually reduced and internally fixated with a 2.4 distal radius locking compression plate. 2D (anterior-posterior and lateral) and 3D (rotational) fluoroscopic images were taken as well as computed tomographies. Fluoroscopic images, Cone Beam CT (CBCT), 360° rotating sequences (so called "Movies") and CT scans were co-evaluated by a specialist orthopedic surgeon and a specialist radiologist regarding quality of fracture reduction, position of plate, position of the three distal locking screws and position of the three diaphyseal screws. In reference to gold standard CT, sensitivity and specifity were analyzed.

RESULTS:

"Movie" showed highest sensitivity for detection of insufficient fracture reduction (88%). Sensitivity for detection of incorrect position of plate was 100% for CBCT and 90% for "Movie." For intraarticular position of screws, 2D fluoroscopy and CBCT showed highest sensitivity and specifity (100 and 91%, respectively). Regarding detection of only marginal intraarticular position of screws, sensitivity and specifity of 2D fluoroscopy reached 100% (CBCT 100 and 83%). "Movie" showed highest sensitivity for detection of overlapping position of screws (100%). When it comes to specifity, CBCT achieved 100%. Regarding detection of only marginal overlapping position of screws, 2D fluoroscopy and "Movie" showed highest sensitivity (100%). CBCT achieved highest specifity (100%).

CONCLUSION:

As for assessment of quality of fracture reduction and detection of incorrect position of plate as well as overlapping position of the three diaphyseal screws CBCT and "Movie" are comparable to CT - especially when combined. Particularly sensitivity is high compared to standard 2D fluoroscopy.
Assuntos
Palavras-chave

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Fraturas do Rádio Limite: Humans Idioma: En Revista: BMC Musculoskelet Disord Assunto da revista: FISIOLOGIA / ORTOPEDIA Ano de publicação: 2021 Tipo de documento: Article País de afiliação: Alemanha

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Fraturas do Rádio Limite: Humans Idioma: En Revista: BMC Musculoskelet Disord Assunto da revista: FISIOLOGIA / ORTOPEDIA Ano de publicação: 2021 Tipo de documento: Article País de afiliação: Alemanha