Detection of Significant Hyperopia in Preschool Children Using Two Automated Vision Screeners.
Maguire, Maureen G; Ying, Gui-Shuang; Ciner, Elise B; Kulp, Marjean Taylor; Candy, T Rowan; Moore, Bruce.
Afiliação
  • Ying GS; Perelman School of Medicine, University of Pennsylvania, Philadelphia, Pennsylvania.
  • Ciner EB; Pennsylvania College of Optometry at Salus University, Elkins Park, Pennsylvania.
  • Kulp MT; The Ohio State University, College of Optometry, Columbus, Ohio.
  • Candy TR; Indiana University School of Optometry, Bloomington, Indiana.
  • Moore B; New England College of Optometry, Boston, Massachusetts.
Optom Vis Sci ; 99(2): 114-120, 2022 02 01.
Article em En | MEDLINE | ID: mdl-34889862
SIGNIFICANCE: Moderate to high uncorrected hyperopia in preschool children is associated with amblyopia, strabismus, reduced visual function, and reduced literacy. Detecting significant hyperopia during screening is important to allow children to be followed for development of amblyopia or strabismus and implementation of any needed ophthalmic or educational interventions. PURPOSE: This study aimed to compare the sensitivity and specificity of two automated screening devices to identify preschool children with moderate to high hyperopia. METHODS: Children in the Vision in Preschoolers (VIP) study were screened with the Retinomax Autorefractor (Nikon, Inc., Melville, NY) and Plusoptix Power Refractor II (Plusoptix, Nuremberg, Germany) and examined by masked eye care professionals to detect the targeted conditions of amblyopia, strabismus, or significant refractive error, and reduced visual acuity. Significant hyperopia (American Association for Pediatric Ophthalmology and Strabismus definition of hyperopia as an amblyopia risk factor), based on cycloplegic retinoscopy, was >4.00 D for age 36 to 48 months and >3.50 D for age older than 48 months. Referral criteria from VIP for each device and from a distributor (PediaVision) for the Power Refractor II were applied to screening results. RESULTS: Among 1430 children, 132 children had significant hyperopia in at least one eye. Using the VIP referral criteria, sensitivities for significant hyperopia were 80.3% for the Retinomax and 69.7% for the Power Refractor II (difference, 10.6%; 95% confidence interval, 7.0 to 20.5%; P = .04); specificities relative to any targeted condition were 89.9 and 89.1%, respectively. Using the PediaVision referral criteria for the Power Refractor, sensitivity for significant hyperopia was 84.9%; however, specificity relative to any targeted condition was 78.3%, 11.6% lower than the specificity for the Retinomax. Analyses using the VIP definition of significant hyperopia yielded results similar to when the American Association for Pediatric Ophthalmology and Strabismus definition was used. DISCUSSION: When implementing vision screening programs for preschool children, the potential for automated devices that use eccentric photorefraction to either miss detecting significant hyperopia or increase false-positive referrals must be taken into consideration.
Assuntos

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Erros de Refração / Seleção Visual / Ambliopia / Estrabismo / Hiperopia Tipo de estudo: Diagnostic_studies / Prognostic_studies / Risk_factors_studies Limite: Child, preschool / Humans Idioma: En Revista: Optom Vis Sci Assunto da revista: OPTOMETRIA Ano de publicação: 2022 Tipo de documento: Article

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Erros de Refração / Seleção Visual / Ambliopia / Estrabismo / Hiperopia Tipo de estudo: Diagnostic_studies / Prognostic_studies / Risk_factors_studies Limite: Child, preschool / Humans Idioma: En Revista: Optom Vis Sci Assunto da revista: OPTOMETRIA Ano de publicação: 2022 Tipo de documento: Article