Evaluation of silicon strip detectors in transmission mode for online beam monitoring in microbeam radiation therapy at the Australian Synchrotron.
Davis, Jeremy; Dipuglia, Andrew; Cameron, Matthew; Paino, Jason; Cullen, Ashley; Guatelli, Susanna; Petasecca, Marco; Rosenfeld, Anatoly; Lerch, Michael.
Afiliação
  • Davis J; Centre for Medical Radiation Physics, University of Wollongong, Wollongong, New South Wales, Australia.
  • Dipuglia A; Centre for Medical Radiation Physics, University of Wollongong, Wollongong, New South Wales, Australia.
  • Cameron M; Centre for Medical Radiation Physics, University of Wollongong, Wollongong, New South Wales, Australia.
  • Paino J; Centre for Medical Radiation Physics, University of Wollongong, Wollongong, New South Wales, Australia.
  • Cullen A; Centre for Medical Radiation Physics, University of Wollongong, Wollongong, New South Wales, Australia.
  • Guatelli S; Centre for Medical Radiation Physics, University of Wollongong, Wollongong, New South Wales, Australia.
  • Petasecca M; Centre for Medical Radiation Physics, University of Wollongong, Wollongong, New South Wales, Australia.
  • Rosenfeld A; Centre for Medical Radiation Physics, University of Wollongong, Wollongong, New South Wales, Australia.
  • Lerch M; Centre for Medical Radiation Physics, University of Wollongong, Wollongong, New South Wales, Australia.
J Synchrotron Radiat ; 29(Pt 1): 125-137, 2022 Jan 01.
Article em En | MEDLINE | ID: mdl-34985430
ABSTRACT
Successful transition of synchrotron-based microbeam radiation therapy (MRT) from pre-clinical animal studies to human trials is dependent upon ensuring that there are sufficient and adequate measures in place for quality assurance purposes. Transmission detectors provide researchers and clinicians with a real-time quality assurance and beam-monitoring instrument to ensure safe and accurate dose delivery. In this work, the effect of transmission detectors of different thicknesses (10 and 375 µm) upon the photon energy spectra and dose deposition of spatially fractionated synchrotron radiation is quantified experimentally and by means of a dedicated Geant4 simulation study. The simulation and experimental results confirm that the presence of the 375 µm thick transmission detector results in an approximately 1-6% decrease in broad-beam and microbeam peak dose. The capability to account for the reduction in dose and change to the peak-to-valley dose ratio justifies the use of transmission detectors as thick as 375 µm in MRT provided that treatment planning systems are able to account for their presence. The simulation and experimental results confirm that the presence of the 10 µm thick transmission detector shows a negligible impact (<0.5%) on the photon energy spectra, dose delivery and microbeam structure for both broad-beam and microbeam cases. Whilst the use of 375 µm thick detectors would certainly be appropriate, based upon the idea of best practice the authors recommend that 10 µm thick transmission detectors of this sort be utilized as a real-time quality assurance and beam-monitoring tool during MRT.
Assuntos
Palavras-chave

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Silício / Síncrotrons Tipo de estudo: Guideline / Health_economic_evaluation Limite: Animals / Humans País/Região como assunto: Oceania Idioma: En Revista: J Synchrotron Radiat Assunto da revista: RADIOLOGIA Ano de publicação: 2022 Tipo de documento: Article País de afiliação: Austrália

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Silício / Síncrotrons Tipo de estudo: Guideline / Health_economic_evaluation Limite: Animals / Humans País/Região como assunto: Oceania Idioma: En Revista: J Synchrotron Radiat Assunto da revista: RADIOLOGIA Ano de publicação: 2022 Tipo de documento: Article País de afiliação: Austrália