Your browser doesn't support javascript.
loading
Grid-mapping catheters versus PentaRay catheters for left atrial mapping on ensite precision mapping system.
Saito, Jumpei; Yamashita, Kennosuke; Numajiri, Takaki; Gibo, Yuma; Usumoto, Soichiro; Narui, Syuro; Fujioka, Tatsuki; Asukai, Yu; Igawa, Wataru; Ono, Morio; Ebara, Seitaro; Okabe, Toshitaka; Isomura, Naoei; Ochiai, Masahiko.
Afiliação
  • Saito J; Division of Cadiology, Northern Yokohama Hospital, Showa University, Yokohama, Japan.
  • Yamashita K; Department of Cardiovascular Medicine, Sendai Kousei Hospital, Sendai, Japan.
  • Numajiri T; Cardiac Arrhythmia Center, Sendai Kousei Hospital, Sendai, Japan.
  • Gibo Y; Division of Cadiology, Northern Yokohama Hospital, Showa University, Yokohama, Japan.
  • Usumoto S; Division of Cadiology, Northern Yokohama Hospital, Showa University, Yokohama, Japan.
  • Narui S; Division of Cadiology, Northern Yokohama Hospital, Showa University, Yokohama, Japan.
  • Fujioka T; Division of Cadiology, Northern Yokohama Hospital, Showa University, Yokohama, Japan.
  • Asukai Y; Division of Cadiology, Northern Yokohama Hospital, Showa University, Yokohama, Japan.
  • Igawa W; Division of Cadiology, Northern Yokohama Hospital, Showa University, Yokohama, Japan.
  • Ono M; Division of Cadiology, Northern Yokohama Hospital, Showa University, Yokohama, Japan.
  • Ebara S; Division of Cadiology, Northern Yokohama Hospital, Showa University, Yokohama, Japan.
  • Okabe T; Division of Cadiology, Northern Yokohama Hospital, Showa University, Yokohama, Japan.
  • Isomura N; Division of Cadiology, Northern Yokohama Hospital, Showa University, Yokohama, Japan.
  • Ochiai M; Division of Cadiology, Northern Yokohama Hospital, Showa University, Yokohama, Japan.
J Cardiovasc Electrophysiol ; 33(7): 1405-1411, 2022 07.
Article em En | MEDLINE | ID: mdl-35441420
ABSTRACT

INTRODUCTION:

Areas displaying reduced bipolar voltage are defined as low-voltage areas (LVAs). Moreover, left atrial (LA) LVAs after pulmonary vein isolation (PVI) have been reported as a predictor of recurrent atrial fibrillation (AF). In this study, we compared grid mapping catheter (GMC) with PentaRay catheter (PC) for LA voltage mapping on Ensite Precision mapping system.

METHODS:

Twenty-six consecutive patients with LVAs and border zone within the LA were enrolled. After achieving PVI, voltage mapping under high right atrial pacing for 600 ms was performed twice using each catheter type (GMC first, PC next). Furthermore, LVA was defined as a region with a bipolar voltage of <0.50, and border zone was defined as a region with a bipolar voltage of <1.0, or <1.5 mV.

RESULTS:

Compared with PC, using GMC, voltage mapping contained more mapping points (20 242 [15 859, 26 013] vs. 5589 [4088, 7649]; p < .0001), and more mapping points per minute(1428 [1275, 1803] vs. 558 [372, 783]; p < .0001). In addition, LVA and border zone size using GMC was significantly less than that reported using PC <1.0 mV (5.9 cm2 [2.9, 20.2] vs. 13.9 cm2 [6.3, 24.1], p = .018) and <1.5 mV voltage cutoff (10.6 cm2 [6.6, 27.2] vs. 21.6 cm2 [12.6, 35.0], p = .005).

CONCLUSION:

Bipolar voltage amplitude estimated by GMC was significantly larger than that estimated by PC on Ensite Precision mapping system. GMC may be able to find highly selective identification of LVAs with lower prevalence and smaller LVA and border zone size.
Assuntos
Palavras-chave

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Veias Pulmonares / Fibrilação Atrial / Ablação por Cateter Tipo de estudo: Diagnostic_studies / Prognostic_studies / Risk_factors_studies Limite: Humans Idioma: En Revista: J Cardiovasc Electrophysiol Assunto da revista: ANGIOLOGIA / CARDIOLOGIA / FISIOLOGIA Ano de publicação: 2022 Tipo de documento: Article País de afiliação: Japão

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Veias Pulmonares / Fibrilação Atrial / Ablação por Cateter Tipo de estudo: Diagnostic_studies / Prognostic_studies / Risk_factors_studies Limite: Humans Idioma: En Revista: J Cardiovasc Electrophysiol Assunto da revista: ANGIOLOGIA / CARDIOLOGIA / FISIOLOGIA Ano de publicação: 2022 Tipo de documento: Article País de afiliação: Japão