Your browser doesn't support javascript.
loading
Computational Photon Counting Using Multithreshold Peak Detection for Fast Fluorescence Lifetime Imaging Microscopy.
Sorrells, Janet E; Iyer, Rishyashring R; Yang, Lingxiao; Martin, Elisabeth M; Wang, Geng; Tu, Haohua; Marjanovic, Marina; Boppart, Stephen A.
Afiliação
  • Sorrells JE; Department of Bioengineering, University of Illinois at Urbana-Champaign, Urbana, Illinois 61801, United States.
  • Iyer RR; Beckman Institute for Advanced Science and Technology, University of Illinois at Urbana-Champaign, Urbana, Illinois 61801, United States.
  • Yang L; Beckman Institute for Advanced Science and Technology, University of Illinois at Urbana-Champaign, Urbana, Illinois 61801, United States.
  • Martin EM; Department of Electrical and Computer Engineering, University of Illinois at Urbana-Champaign, Urbana, Illinois 61801, United States.
  • Wang G; Beckman Institute for Advanced Science and Technology, University of Illinois at Urbana-Champaign, Urbana, Illinois 61801, United States.
  • Tu H; Department of Electrical and Computer Engineering, University of Illinois at Urbana-Champaign, Urbana, Illinois 61801, United States.
  • Marjanovic M; Department of Bioengineering, University of Illinois at Urbana-Champaign, Urbana, Illinois 61801, United States.
  • Boppart SA; Beckman Institute for Advanced Science and Technology, University of Illinois at Urbana-Champaign, Urbana, Illinois 61801, United States.
ACS Photonics ; 9(8): 2748-2755, 2022 Aug 17.
Article em En | MEDLINE | ID: mdl-35996369
ABSTRACT
Time-resolved photon counting methods have a finite bandwidth that restricts the acquisition speed of techniques like fluorescence lifetime imaging microscopy (FLIM). To enable faster imaging, computational methods can be employed to count photons when the output of a detector is directly digitized at a high sampling rate. Here, we present computational photon counting using a hybrid photodetector in conjunction with multithreshold peak detection to count instances where one or more photons arrive at the detector within the detector response time. This method can be used to distinguish up to five photon counts per digitized point, whereas previous demonstrations of computational photon counting on data acquired with photomultiplier tubes have only counted one photon at a time. We demonstrate in both freely moving C. elegans and a human breast cancer cell line undergoing apoptosis that this novel multithreshold peak detection method can accurately characterize the intensity and fluorescence lifetime of samples producing photon rates up to 223%, higher than previously demonstrated photon counting FLIM systems.

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Tipo de estudo: Diagnostic_studies Idioma: En Revista: ACS Photonics Ano de publicação: 2022 Tipo de documento: Article País de afiliação: Estados Unidos

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Tipo de estudo: Diagnostic_studies Idioma: En Revista: ACS Photonics Ano de publicação: 2022 Tipo de documento: Article País de afiliação: Estados Unidos