Your browser doesn't support javascript.
loading
[Clinical study on application of 3D Slicer software assisted domestic frameless stereotactic robot in biopsy of intracranial lesions].
Chen, H; Yan, X; He, F; Ding, S C; Diao, J F; Guo, H; Cao, S M; Yang, C J; Yin, F.
Afiliação
  • Chen H; Department of Neurosurgery, Aerospace Central Hospital, Beijing 100049, China.
  • Yan X; Department of Neurosurgery, Aerospace Central Hospital, Beijing 100049, China.
  • He F; Department of Neurosurgery, Aerospace Central Hospital, Beijing 100049, China.
  • Ding SC; Department of Neurosurgery, Aerospace Central Hospital, Beijing 100049, China.
  • Diao JF; Department of Neurosurgery, Aerospace Central Hospital, Beijing 100049, China.
  • Guo H; Department of Neurosurgery, Aerospace Central Hospital, Beijing 100049, China.
  • Cao SM; Department of Neurosurgery, Aerospace Central Hospital, Beijing 100049, China.
  • Yang CJ; Department of Neurosurgery, Aerospace Central Hospital, Beijing 100049, China.
  • Yin F; Department of Neurosurgery, Aerospace Central Hospital, Beijing 100049, China.
Zhonghua Wai Ke Za Zhi ; 61(1): 61-65, 2023 Jan 01.
Article em Zh | MEDLINE | ID: mdl-36603886
ABSTRACT

Objective:

To examine the application value of 3D Slicer software assisted domestic frameless stereotactic robot in biopsy of intracranial lesions.

Methods:

A retrospective analysis was performed on 80 patients who admitted consecutively and underwent intracerebral lesions biopsy with the domestic frameless stereotactic robot at Department of Neurosurgery, Aerospace Central Hospital from January 2019 to December 2021. There were 36 males and 44 females, with a mean age of (38.5±18.0) years (range 6 to 71 years). Before surgery only enhanced T1-weighted three-dimensional magnetization prepared gradient echo sequences and diffusion tensor imaging scans were performed. Self-reconstruction of intracranial lesions, cerebral cortex and blood vessels was carried out using 3D Slicer software system after the DICOM format imaging data of 80 patients were collected. These imaging data were merged to the workstation of the domestic frameless stereotactic robot for preoperative surgical planning and the surgical puncture path was designed to avoid blood vessels in the brain functional area, cerebral cortex and sulcus.

Results:

All frameless stereotactic biopsy were successfully performed. Postoperative pathological diagnosis included 50 cases of diffuse astrocytic and oligodendroglioma, 15 cases of lymphoma, 5 cases of metastatic tumors, 5 cases of inflammatory demyelinating disease, 2 cases of inflammatory granuloma, 1 case of hemangioma, 1 case of acute lymphoblastic leukemia intracranial invasion and 1 case of seminoma. The positive diagnosis rate was 100% (80/80). Postoperative imaging confirmed that the puncture path and target were accurately implemented according to the preoperative planning, and the target error was (1.32±0.44) mm (range 0.55 to 1.99 mm). One case of puncture-related bleeding occurred at the target after surgery and improved after treatment.

Conclusion:

The three-dimensional multimodal images reconstructed by the 3D Slicer software before operation could help the surgeons make the preoperative planning and reduce the risk of stereotactic brain biopsy.
Assuntos

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Robótica / Neoplasias Encefálicas Tipo de estudo: Observational_studies / Risk_factors_studies Limite: Adolescent / Adult / Aged / Child / Female / Humans / Male / Middle aged Idioma: Zh Revista: Zhonghua Wai Ke Za Zhi Ano de publicação: 2023 Tipo de documento: Article País de afiliação: China

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Robótica / Neoplasias Encefálicas Tipo de estudo: Observational_studies / Risk_factors_studies Limite: Adolescent / Adult / Aged / Child / Female / Humans / Male / Middle aged Idioma: Zh Revista: Zhonghua Wai Ke Za Zhi Ano de publicação: 2023 Tipo de documento: Article País de afiliação: China