Your browser doesn't support javascript.
loading
Determination of serum CA724 levels using fluorescence immunochromatography.
Liu, Chang; Chen, Cuicui; Peng, Fenglan; Liang, Huankun; Zhong, Shuhai; Liu, Tiancai; Li, Laiqing; Boronin, Alexander; Dong, Wenqi.
Afiliação
  • Liu C; Heilongjiang Academy of Chinese Medicine Sciences, Harbin, 150036, China.
  • Chen C; Guangzhou Youdi Bio-technology Co., Ltd, Guangzhou, 510663, China.
  • Peng F; Jinan Laide Bio-technology Co., Ltd, Jinan, 271100, China.
  • Liang H; Taian Disabled Soldiers' Hospital of Shandong Province, Taian, 271000, China.
  • Zhong S; Guangzhou Youdi Bio-technology Co., Ltd, Guangzhou, 510663, China.
  • Liu T; Guangzhou Youdi Bio-technology Co., Ltd, Guangzhou, 510663, China.
  • Li L; School of Laboratory Medicine and Biotechnology, Southern Medical University, Guangzhou, 510515, China.
  • Boronin A; Guangzhou Youdi Bio-technology Co., Ltd, Guangzhou, 510663, China.
  • Dong W; Laboratory of Plasmid Biology, G.K. Skryabin Institute Of Biochemistry and Physiology of Microorganisms Russian Academy of Sciences, Pushchino, 142290, Russia.
BMC Biotechnol ; 23(1): 30, 2023 08 18.
Article em En | MEDLINE | ID: mdl-37596558
ABSTRACT

BACKGROUND:

Carbohydrate antigen 724 (CA724) is a sensitive and specific indicator for multiple malignant tumors. The aim of this study was to establish a Eu-time resolved fluorescence immunochromatography (Eu-TRFICO) method for quantitative detection of CA724 in serum.

METHODS:

Eu-TRFICO strips were optimized and assembled. The sensitivity, specificity and precision were evaluated using CA724 standard dilutions and matrix serum. Meanwhile, the reference interval, comparison, and sensitivity/specificity were performed using clinical negative/positive gastric cancer serum samples.

RESULTS:

The standard curve equation was y = 9.869 x - 154.12 (R2 = 0.993), and the sensitivity was 0.42 U/mL. The common interferents in serum could not affect the quantitative results with low cross-reactivities (all no more than 1.09%). All average recoveries of the intra- and interbatch ranged from 102.38 to 106.40%, and all CVs were below 10%. The reference interval of the healthy subjects was < 4.68 U/mL and the reference interval of the subjects with grade I/II gastric cancer was > 9.54 U/mL. Additionally, a high Pearson r (0.9503) and sensitivity/specificity (92.86%/94.20%) were obtained.

CONCLUSION:

This study prepared Eu-TRFICO strips with high sensitivity, specificity, precision and satisfactory clinical testing performance, which provides more options for clinical quantitative and convenient testing of CA724.
Assuntos
Palavras-chave

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Neoplasias Gástricas Tipo de estudo: Diagnostic_studies Limite: Humans Idioma: En Revista: BMC Biotechnol Assunto da revista: BIOTECNOLOGIA Ano de publicação: 2023 Tipo de documento: Article País de afiliação: China

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Neoplasias Gástricas Tipo de estudo: Diagnostic_studies Limite: Humans Idioma: En Revista: BMC Biotechnol Assunto da revista: BIOTECNOLOGIA Ano de publicação: 2023 Tipo de documento: Article País de afiliação: China