Simplified assessment for chemical exchanged saturation transfer (CEST) imaging: local offset frequency and CEST effect.
Chiba, Daiki; Kanazawa, Yuki; Miyati, Tosiaki; Harada, Masafumi; Miyoshi, Mitsuharu; Hayashi, Hiroaki; Haga, Akihiro.
Afiliação
  • Chiba D; Brain Attack Center Ota Memorial Hospital, 3-6-28 Okinogami, Fukuyama, Hiroshima, 7200825, Japan.
  • Kanazawa Y; Institute of Biomedical Sciences, Tokushima University Graduate School, 3-18-15 Kuramoto-Cyo, Toksuhima, 7708503, Japan. yk@tokushima-u.ac.jp.
  • Miyati T; Faculty of Health Sciences, Institute of Medical, Pharmaceutical and Health Sciences, Kanazawa University, 5­11­80 Kodatsuno, Kanazawa, Ishikawa, 9200942, Japan.
  • Harada M; Institute of Biomedical Sciences, Tokushima University Graduate School, 3-18-15 Kuramoto-Cyo, Toksuhima, 7708503, Japan.
  • Miyoshi M; Global MR Clinical Solutions and Research Collaborations, GE HealthCare, 4-7-127 Asahigaoka, Hino, Tokyo, 1918503, Japan.
  • Hayashi H; Faculty of Health Sciences, Institute of Medical, Pharmaceutical and Health Sciences, Kanazawa University, 5­11­80 Kodatsuno, Kanazawa, Ishikawa, 9200942, Japan.
  • Haga A; Institute of Biomedical Sciences, Tokushima University Graduate School, 3-18-15 Kuramoto-Cyo, Toksuhima, 7708503, Japan.
Radiol Phys Technol ; 17(1): 93-102, 2024 Mar.
Article em En | MEDLINE | ID: mdl-37897684
The aim of this study is to develop a novel phantom for the evaluation of clinical CEST imaging settings, e.g., B0 and B1 field inhomogeneities, CEST contrast, and post-processing. We made a phantom composed of two slice sections: a grid section for local offset frequency evaluation and a sample section for CEST effect evaluation using different concentrations of an egg white albumin solution. On a 3 Tesla MR scanner, a phantom study was performed using CEST imaging; the mean B1 amplitudes were set at 1.2 and 1.9 µT, and CEST images with and without B0 corrections were acquired. Next, region of interest (ROI) analysis was performed for each slice. Then, CEST images with and without B0 corrections were compared at each B1 amplitude. The B0 corrected Z-spectrums at each local region in the grid section showed a shifting of the curve bottom to 0 ppm. Z-spectrum at B1 = 1.9 µT showed a broader curve shape than that at 1.2 µT. Moreover, MTRasym values at 3.5 ppm for each albumin sample at B1 = 1.9 µT were about two times higher than those at 1.2 µT. Our phantom enabled us to evaluate and optimize B0 inhomogeneity and the CEST effect at the B1 amplitude.