Application of the cell-based RT-qPCR assay (C-QPA) for potency detection of the novel trivalent rotavirus vaccine in China.
Wang, Yunjin; Liu, YueYue; Bao, Hong; Chen, Yueru; Kou, Guiying; Wang, Mingqiang; Fu, Shengfang; Huo, Wen; Guan, Wenzhu; Cheng, Yahui; Zhou, Xu; Li, Xiongxiong.
Afiliação
  • Wang Y; Lanzhou Institute of Biological Products Co., Ltd., Lanzhou, China.
  • Liu Y; National Institutes for Food and Drug Control, Beijing, China.
  • Bao H; Lanzhou Institute of Biological Products Co., Ltd., Lanzhou, China.
  • Chen Y; Lanzhou Institute of Biological Products Co., Ltd., Lanzhou, China.
  • Kou G; Lanzhou Institute of Biological Products Co., Ltd., Lanzhou, China.
  • Wang M; Lanzhou Institute of Biological Products Co., Ltd., Lanzhou, China.
  • Fu S; Lanzhou Institute of Biological Products Co., Ltd., Lanzhou, China.
  • Huo W; Lanzhou Institute of Biological Products Co., Ltd., Lanzhou, China.
  • Guan W; Lanzhou Institute of Biological Products Co., Ltd., Lanzhou, China.
  • Cheng Y; Lanzhou Institute of Biological Products Co., Ltd., Lanzhou, China.
  • Zhou X; Shanghai Institute of Biological Products Co., Ltd., Shanghai, China.
  • Li X; Lanzhou Institute of Biological Products Co., Ltd., Lanzhou, China.
J Clin Lab Anal ; 37(23-24): e24989, 2023 Dec.
Article em En | MEDLINE | ID: mdl-37975330
ABSTRACT

BACKGROUND:

Because of the deficiencies of traditional methods in multivalent rotavirus vaccine potency detection, a cell-based quantitative RT-qPCR assay (C-QPA) was established and validated for specificity, precision, and accuracy.

METHODS:

In order to further validate the robustness of this method in actual titer detection, the linear range and the practical application under different conditions were tested using monovalent and trivalent rotavirus samples and standards.

RESULTS:

Results showed that the linear range was 2.0-6.5, 3.9-8.3, and 3.5-8.1 UI (unit of infectivity) for G2, G3, and G4, respectively. Besides, unknown sample with high titer exceeding the linear range can be calculated by dilution. The UIs of serotypes G2, G3, and G4 in monovalent and trivalent rotavirus samples showed a relative deviation ≤4.10%, and the monovalent samples of the same serotype with or without protective agents showed a relative deviation ≤4.28%; the coefficient of variation (CV) of at least 176 tests (548 individual runs) of 3 in vitro-transcribed RNA standards with certain concentrations was not higher than 6.50%; the results of the trivalent samples tested by more than 149 times in 5 years (467 individual runs) showed the CVs lower than 12.66%; 15 samples detected by one laboratory showed a CV lower than 9.83%, while other three samples tested by two independent laboratories showed a CV lower than 6.90%.

CONCLUSION:

In summary, the C-QPA has good linearity, durability, repeatability, and reproducibility in practical application and has been proved by the authority to be widely used in the production, quality control and release of the recently licensed trivalent vaccine in China.
Assuntos
Palavras-chave

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Infecções por Rotavirus / Rotavirus / Vacinas contra Rotavirus Limite: Humans País/Região como assunto: Asia Idioma: En Revista: J Clin Lab Anal Assunto da revista: TECNICAS E PROCEDIMENTOS DE LABORATORIO Ano de publicação: 2023 Tipo de documento: Article País de afiliação: China

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Infecções por Rotavirus / Rotavirus / Vacinas contra Rotavirus Limite: Humans País/Região como assunto: Asia Idioma: En Revista: J Clin Lab Anal Assunto da revista: TECNICAS E PROCEDIMENTOS DE LABORATORIO Ano de publicação: 2023 Tipo de documento: Article País de afiliação: China