ABSTRACT
Biological monitoring of the unmodified sevoflurane and its metabolite hexafluoroisopropanol (HFIP) in urine samples was proposed to determine the individual exposure levels of the medical staff. In this study, a method for simultaneous determination of both compounds in urine using static headspace-gas chromatography-mass spectrometry (HS-GC-MS) was developed. The method is linear over a broad concentration range from 1 to 1000 µg/L (r2 > 0.999) and shows high precision. Limits of quantification (LOQ) are 0.6 µg/L for sevoflurane and 3 µg/L for HFIP, representing an excellent sensitivity without the necessity of analyte enrichment. The method was successfully applied in a German pilot-study to monitor both compounds in samples from medical personnel working in operating theatres. Urinary concentrations of HFIP ranged between < LOQ and 145 µg/L, while sevoflurane was below the LOD in all samples.