Your browser doesn't support javascript.
loading
Assessment of measurement accuracy of amplified DNA using a colorimetric loop-mediated isothermal amplification assay.
Hong, Seongsoo; Park, Jeongho; Bok, Jaekyung; Cho, Euna; Rhee, Joowon.
Afiliação
  • Hong S; 1drop Inc, 215, Galmachi-ro, Jungwon-gu, Seongnam-si, Gyeonggi-do, Korea.
  • Park J; 1drop Inc, 215, Galmachi-ro, Jungwon-gu, Seongnam-si, Gyeonggi-do, Korea.
  • Bok J; 1drop Inc, 215, Galmachi-ro, Jungwon-gu, Seongnam-si, Gyeonggi-do, Korea.
  • Cho E; 1drop Inc, 215, Galmachi-ro, Jungwon-gu, Seongnam-si, Gyeonggi-do, Korea.
  • Rhee J; 1drop Inc, 215, Galmachi-ro, Jungwon-gu, Seongnam-si, Gyeonggi-do, Korea.
Biotechniques ; 76(3): 114-118, 2024 Mar.
Article em En | MEDLINE | ID: mdl-38131320
ABSTRACT
A colorimetric loop-mediated isothermal amplification assay detects changes in pH during amplification based on color changes at a constant temperature. Currently, various studies have focused on developing and assessing molecular point-of-care testing instruments. In this study, we evaluated amplified DNA concentrations measured using the colorimetric LAMP assay of the 1POT™ Professional device (1drop Inc, Korea). Results of the 1POT analysis of clinical samples were compared with measurements obtained from the Qubit™ 4 and NanoDrop™ 2000 devices (both from Thermo Fisher Scientific, MA, USA). These results showed a correlation of 0.98 (95% CI 0.96-0.99) and 0.96 (95% CI 0.92-0.98) between 1POT and the Qubit and NanoDrop. 1POT can measure amplified DNA accurately and is suitable for on-site molecular diagnostics.
Assuntos
Palavras-chave

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: DNA / Colorimetria Idioma: En Revista: Biotechniques Ano de publicação: 2024 Tipo de documento: Article