Your browser doesn't support javascript.
loading
Tube obstruction caused by intraocular lens capture following PreserFlo MicroShunt implantation.
Iwasaki, Kentaro; Arimura, Shogo; Inatani, Masaru.
Afiliação
  • Iwasaki K; Department of Ophthalmology, Faculty of Medical Sciences, University of Fukui, Fukui, Japan.
  • Arimura S; Department of Ophthalmology, Faculty of Medical Sciences, University of Fukui, Fukui, Japan.
  • Inatani M; Department of Ophthalmology, Faculty of Medical Sciences, University of Fukui, Fukui, Japan.
Am J Ophthalmol Case Rep ; 33: 101951, 2024 Mar.
Article em En | MEDLINE | ID: mdl-38192607
ABSTRACT

Purpose:

We describe a rare case of tube obstruction caused by intraocular lens (IOL) capture following a PreserFlo MicroShunt (PMS) surgery and its subsequent management. Observations Tube obstruction was noted following PMS implantation at 8 days postoperatively. The intraocular pressure (IOP) increased to 42 mmHg because of tube occlusion that was caused by iris and IOL capture at the tip of the tube. The occlusion was released surgically to free the lumen, and the IOP rapidly decreased to 14 mmHg. Conclusions and importance IOP elevation due to tube obstruction caused by iris and IOL capture after PMS surgery was resolved by surgical intervention without tube reinsertion. Extra care is required regarding the IOL position in relation to the PMS tube when hypotony occurs in the early postoperative period.
Palavras-chave

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Idioma: En Revista: Am J Ophthalmol Case Rep Ano de publicação: 2024 Tipo de documento: Article País de afiliação: Japão