Nonuniform sliding-window reconstruction for accelerated dual contrast agent quantification with MR fingerprinting.
Marriott, Anna; Rioux, James; Brewer, Kimberly.
Afiliação
  • Marriott A; Biomedical MRI Research Laboratory (BMRL), IWK Health Centre, 5850/5980 University Avenue, Halifax, NS, B3K 6R8, Canada.
  • Rioux J; Department of Physics and Atmospheric Science, Dalhousie University, Halifax, NS, Canada.
  • Brewer K; Department of Physics and Atmospheric Science, Dalhousie University, Halifax, NS, Canada.
MAGMA ; 37(2): 273-282, 2024 Apr.
Article em En | MEDLINE | ID: mdl-38217784
ABSTRACT

OBJECTIVE:

MR fingerprinting (MRF) can enable preclinical studies of cell tracking by quantifying multiple contrast agents simultaneously, but faster scan times are required for in vivo applications. Sliding window (SW)-MRF is one option for accelerating MRF, but standard implementations are not sufficient to preserve the accuracy of T2*, which is critical for tracking iron-labelled cells in vivo.

PURPOSE:

To develop a SW approach to MRF which preserves the T2* accuracy required for accelerated concentration mapping of iron-labelled cells on single-channel preclinical systems.

METHODS:

A nonuniform SW was applied to the MRF sequence and dictionary. Segments of the sequence most sensitive to T2* were subject to a shorter window length, preserving the T2* sensitivity. Phantoms containing iron-labelled CD8+ T cells and gadolinium were used to compare 24× undersampled uniform and nonuniform SW-MRF parameter maps. Dual concentration maps were generated for both uniform and nonuniform MRF and compared.

RESULTS:

Lin's concordance correlation coefficient, compared to gold standard parameter values, was much greater for nonuniform SW-MRF than for uniform SW-MRF. A Wilcoxon signed-rank test showed no significant difference between nonuniform SW-MRF and gold standards. Nonuniform SW-MRF outperformed the uniform SW-MRF concentration maps for all parameters, providing a balance between T2* sensitivity of short window lengths, and SNR of longer window lengths.

CONCLUSIONS:

Nonuniform SW-MRF improves the accuracy of matching compared to uniform SW-MRF, allowing higher accelerated concentration mapping for preclinical systems.