Elevating Platinum to Volumetric Capacitance: High Surface Area Electrodes through Reactive Pt Sputtering.
Adv Healthc Mater
; 13(24): e2302400, 2024 Sep.
Article
em En
| MEDLINE
| ID: mdl-38758352
ABSTRACT
Platinum is the most widespread electrode material used for implantable biomedical and neuroelectronic devices, motivating exploring ways to improve its performance and understand its fundamental properties. Using reactive magnetron sputtering, PtOx is prepared, which upon partial reduction yields a porous thin-film form of platinum with favorable properties, notably record-low impedance values outcompeting other reports for platinum-based electrodes. It is established that its high electrochemical capacitance scales with thickness, in the way of volumetric capacitor materials like IrOx and poly(3,4-ethylenedioxythiophene), PEDOT. Unlike these two well-known analogs, however, it is found that PtOx capacitance is not caused by reversible pseudofaradaic reactions but rather due to high surface area. In contrast to IrOx, PtOx is not a reversible valence-change oxide, but rather a porous form of platinum. The findings show that this oxygen-containing form of Pt can place Pt electrodes on a level competitive with IrOx and PEDOT. Due to its relatively low cost and ease of preparation, PtOx can be a good choice for microfabricated bioelectronic devices.
Palavras-chave
Texto completo:
1
Coleções:
01-internacional
Base de dados:
MEDLINE
Assunto principal:
Platina
/
Compostos Bicíclicos Heterocíclicos com Pontes
/
Capacitância Elétrica
/
Eletrodos
Idioma:
En
Revista:
Adv Healthc Mater
Ano de publicação:
2024
Tipo de documento:
Article
País de afiliação:
Suécia