Cranial ultrasound in preterm infants ≤ 32 weeks gestation-novel insights from the use of very high-frequency (18-5 MHz) transducers: a case series.
Miselli, Francesca; Guidotti, Isotta; Di Martino, Marianna; Bedetti, Luca; Minotti, Chiara; Spaggiari, Eugenio; Malmusi, Giovanni; Lugli, Licia; Corso, Lucia; Berardi, Alberto.
Afiliação
  • Miselli F; PhD Program in Clinical and Experimental Medicine, University of Modena and Reggio Emilia, Modena, Italy.
  • Guidotti I; Neonatal Intensive Care Unit, Women's and Children's Health Department, University Hospital of Modena, Via del Pozzo, 41124, Modena, Italy.
  • Di Martino M; Neonatal Intensive Care Unit, Women's and Children's Health Department, University Hospital of Modena, Via del Pozzo, 41124, Modena, Italy.
  • Bedetti L; School of Pediatrics Residency, University of Modena and Reggio Emilia, 41224, Modena, Italy.
  • Minotti C; PhD Program in Clinical and Experimental Medicine, University of Modena and Reggio Emilia, Modena, Italy. luca.bedetti@unimore.it.
  • Spaggiari E; Neonatal Intensive Care Unit, Women's and Children's Health Department, University Hospital of Modena, Via del Pozzo, 41124, Modena, Italy. luca.bedetti@unimore.it.
  • Malmusi G; Neonatal Intensive Care Unit, Women's and Children's Health Department, University Hospital of Modena, Via del Pozzo, 41124, Modena, Italy.
  • Lugli L; Neonatal Intensive Care Unit, Women's and Children's Health Department, University Hospital of Modena, Via del Pozzo, 41124, Modena, Italy.
  • Corso L; Neonatal Intensive Care Unit, Italy Department of Obstetrics and Pediatrics, Azienda USL-IRCCS di Reggio Emilia, Reggio Emilia, Italy.
  • Berardi A; Neonatal Intensive Care Unit, Women's and Children's Health Department, University Hospital of Modena, Via del Pozzo, 41124, Modena, Italy.
Eur J Pediatr ; 183(8): 3589-3598, 2024 Aug.
Article em En | MEDLINE | ID: mdl-38831135
ABSTRACT
The quality of cranial ultrasound has improved over time, with advancing technology leading to higher resolution, faster image processing, digital display, and back-up. However, some brain lesions may remain difficult to characterize since higher frequencies result in greater spatial resolution, the use of additional transducers may overcome some of these limitations. The very high-frequency transducers (18-5 MHz) are currently employed for small parts and lung ultrasound. Here we report the first case series comparing the very high-frequency probes (18-5 MHz) with standard micro-convex probes (8-5 MHz) for cranial ultrasound in preterm infants. In this case series, we compared cranial ultrasound images obtained with a micro-convex transducer (8-5 MHz) and those obtained with a very high-frequency (18-5 MHz) linear array transducer in 13 preterm infants ≤ 32 weeks gestation (9 with cerebral abnormalities and 4 with normal findings). Ultrasound examinations using the very high-frequency linear transducer and the standard medium-frequency micro-convex transducer were performed simultaneously. We also compared ultrasound findings with brain MRI images obtained at term corrected age. Ultrasound images obtained with the very high-frequency (18-5 MHz) transducer showed high quality and accuracy. Notably, despite their higher frequency and expected limited penetration capacity, brain size is small enough in preterm infants, so that brain structures are close to the transducer, allowing for complete evaluation.   

Conclusion:

 We propose the routine use of very high-frequency linear probes as a complementary scanning modality for cranial ultrasound in preterm infants ≤ 32 weeks gestation. What is Known • Brain lesions in preterm infants may remain insufficiently defined through conventional cranial ultrasound scan. • Higher frequency probes  offer better spatial resolution but have a narrower filed of exploration and limited penetration capacity. What is New • Very high-frequency probes were compared with standard medium-frequency probes for cranial ultrasound in infants  ≤ 32 weeks' gestation. • Thanks to the smaller skull size of preterm infants, the new very high-frequency transducers allowed a complete and accurate evaluation.