Comparison of radiation-shielding curtains for endoscopic retrograde cholangiopancreatography staff.
Ishii, Hiroki; Chida, Koichi; Inaba, Yohei; Onodera, Shu; Sai, Masahiro; Zuguchi, Masayuki.
Afiliação
  • Ishii H; Department of Radiological Technology, Tohoku University Hospital, 1-1 Seiryo-machi, Aoba-ku, Sendai 980-8574, Japan.
  • Chida K; Department of Radiological Technology, Tohoku University Graduate School of Medicine, 2-1 Seiryo-machi, Aoba-ku, Sendai 980-8575, Japan.
  • Inaba Y; Department of Radiological Technology, Tohoku University Graduate School of Medicine, 2-1 Seiryo-machi, Aoba-ku, Sendai 980-8575, Japan.
  • Onodera S; Division of Disaster Medical Science, International Research Institute of Disaster Science, Tohoku University, 468-1 Aoba, Aramaki, Aoba-ku, Sendai 980-8572, Japan.
  • Sai M; Department of Radiological Technology, Tohoku University Graduate School of Medicine, 2-1 Seiryo-machi, Aoba-ku, Sendai 980-8575, Japan.
  • Zuguchi M; Division of Disaster Medical Science, International Research Institute of Disaster Science, Tohoku University, 468-1 Aoba, Aramaki, Aoba-ku, Sendai 980-8572, Japan.
J Radiol Prot ; 44(3)2024 Sep 25.
Article em En | MEDLINE | ID: mdl-39265581
ABSTRACT
Occupational radiation exposure to the eye lens of medical staff during endoscopic retrograde cholangiopancreatography (ERCP) should be kept low so as not to exceed annual dose limits. Dose should be low to avoid tissue reactions and minimizing stochastic effects. It is known that the head and neck of the staff are exposed to more scattered radiation in an over-couch tube system than in a C-arm system (under-couch tube). However, this is only true when radiation-shielding curtains are not used. This study aimed to compare the protection radiation to the occupationally exposed worker between a lead curtain mounted on a C-arm system and an ERCP-specific lead curtain mounted on an over-couch tube system. A phantom study simulating a typical setting for ERCP procedures was conducted, and the scattered radiation dose at four staff positions were measured. It was found that scattered radiation doses were higher in the C-arm with a lead curtain than in the over-couch tube with an ERCP-specific lead curtain at all positions measured in this study. It was concluded that the over-couch tube system with an ERCP-specific lead curtain would reduce the staff eye dose by less than one-third compared to the C-arm system with a lead curtain. For the C-arm system, it is necessary to consider more effective radiation protection measures for the upper body of the staff, such as a ceiling-suspended lead screen or another novel shielding that do not interfere with procedures.
Assuntos
Palavras-chave

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Proteção Radiológica / Exposição Ocupacional / Colangiopancreatografia Retrógrada Endoscópica Limite: Humans Idioma: En Revista: J Radiol Prot Assunto da revista: RADIOLOGIA Ano de publicação: 2024 Tipo de documento: Article País de afiliação: Japão

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Proteção Radiológica / Exposição Ocupacional / Colangiopancreatografia Retrógrada Endoscópica Limite: Humans Idioma: En Revista: J Radiol Prot Assunto da revista: RADIOLOGIA Ano de publicação: 2024 Tipo de documento: Article País de afiliação: Japão