ABSTRACT
A new high-performance liquid chromatographic (HPLC) assay without any extraction procedure was developed for the quantification of fluindione in plasma using a 100-microl sample volume and coumarin as the internal standard. A deproteinization procedure was coupled with a reversed-phase HPLC separation using a 250x4.6 mm I.D. C18 column and a UV detector set at 280 nm. Peak height ratios were linear over the range 0.05 to 10 microg/ml (correlation coefficient >0.998). The method was found to be highly reproducible, as indicated by the low value obtained for the coefficient of variation C.V. < or = 6.1% (n = 10). The limit of quantification, estimated under the described conditions at a signal-to-noise ratio of three and with a C.V. lower than 20% for precision and accuracy, was 0.025 microg/ml. The total turnaround time was 25 min. After storage of blood samples at concentrations of 0.1, 0.5 and 2.5 microg/ml at room temperature and exposition to light for 120 h, no degradation of fluindione occurred. This micromethod is simple (no extraction step), fast and currently is being used for drug monitoring.
Assuntos
Buscar no Google
Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Fenindiona / Anticoagulantes Limite: Humans Idioma: En Revista: J Chromatogr B Biomed Sci Appl Assunto da revista: QUIMICA CLINICA Ano de publicação: 1998 Tipo de documento: Article País de afiliação: França
Buscar no Google
Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Fenindiona / Anticoagulantes Limite: Humans Idioma: En Revista: J Chromatogr B Biomed Sci Appl Assunto da revista: QUIMICA CLINICA Ano de publicação: 1998 Tipo de documento: Article País de afiliação: França