Invest Radiol ; 36(11): 677-81, 2001 Nov.
Artículo en Inglés | MEDLINE | ID: mdl-11606845

RESUMEN

RATIONALE AND OBJECTIVES: To optimize the intraluminal signal intensity of a nitinol stent by performing contrast-enhanced three-dimensional magnetic resonance angiography (CE-MRA) with varying flip angles (FAs). METHODS: Contrast-enhanced magnetic resonance angiography at 1.5 T and FAs of 30 degrees, 100 degrees, and 150 degrees was performed on five sheep with 10 iliac nitinol stents (Memotherm-FLEXX). Maximum-intensity projections (MIPs) and composite images of MIPs were performed and compared. RESULTS: Reconstructed MIPs at an FA of 150 degrees showed a slightly disturbed lumen visibility inside the stent accompanied by low-grade lumen visibility outside the stent and vice versa for an FA of 30 degrees. Composite images of a 30 degrees MIP added to a 150 degrees MIP resulted in improved image quality compared with the standard MIP of a single FA. CONCLUSIONS: Signal loss due to radiofrequency shielding inside nitinol stents imaged by CE-MRA can be reduced by applying high FAs. Composite MIP images allow simultaneous visualization of the lumen inside as well as outside the stent.


Asunto(s)
Angiografía por Resonancia Magnética/métodos , Stents , Aleaciones , Animales , Medios de Contraste , Gadolinio DTPA , Arteria Ilíaca/patología , Arteria Ilíaca/cirugía , Masculino , Ovinos