Appl Spectrosc ; 59(4): 442-51, 2005 Apr.
Artículo en Inglés | MEDLINE | ID: mdl-15901329

RESUMEN

The aim of this study was to demonstrate that mid-infrared spectroscopy is able to quantify glucose in a serum matrix with sample volumes well below 1 muL. For this, we applied mid-infrared attenuated total reflectance (ATR) or transmission-based spectroscopic methods to glucose quantification in microsamples of dry-film sera, either undiluted or diluted 10 times in distilled water. The sample series spanned physiological glucose concentrations between 50 and 600 mg/dL and volumes of 80, 8, and 1 nL. Calibration was carried out using multivariate partial least-squares (PLS) modeling with spectral data between 1180 and 940 cm(-1). Best performance was achieved in the ATR experiments. For raw ATR spectra, the optimum standard error of prediction (SEP) of 13.3 mg/dL was obtained for the 8 nL sample series with subsequent 10-fold dilution. With respect to the coefficient of variation of the glucose assay, CV(pred), we obtained a value of 3% for the 80 nL volume samples with spectral preprocessing using matrix protein absorption bands as an internal standard, 4% for the 8 nL samples, and 6% for the 1 nL samples with raw data. Spectral standardization resulted in significant improvement, especially for the 80 nL volume sample series. By contrast, the accuracy of the glucose assay for the 1 nL sample volume series could not be improved either by internal standardization or by considering the dry film areas for normalization, which we attribute to varying topographies of the dry films.


Asunto(s)
Análisis Químico de la Sangre/métodos , Glucemia/análisis , Microquímica/métodos , Nanotecnología/métodos , Refractometría/métodos , Espectroscopía Infrarroja por Transformada de Fourier/métodos , Reproducibilidad de los Resultados , Sensibilidad y Especificidad
2.
Appl Spectrosc ; 58(4): 442-50, 2004 Apr.
Artículo en Inglés | MEDLINE | ID: mdl-15104814

RESUMEN

The aim of this study was to determine the feasibility of minimally invasive glucose concentration measurement of a body fluid within the physiologically important range below 100 nL with a number of samples such as interstitial fluid, plasma, or whole blood using mid-infrared spectroscopy, but starting with preliminary measurements on samples of simple aqueous glucose solutions. The Fourier transform infrared spectrometer was equipped with a Golden Gate single reflection diamond attenuated total reflection (ATR) accessory and a room-temperature pyroelectric detector. As the necessary detection limits can be achieved only for dried samples within the spectrometric conditions realized by a commercial instrument, the work focused on the optimization of such ATR measurements. We achieved quantification of samples with volumes as low as 7 nL between 10 and 600 mg/dL. The standard error of prediction (SEP) for the concentration range 10-100 mg/dL is 3.2 mg/dL with full interval data between 1180 and 940 cm(-1). The performance of the prediction is given by a coefficient of variation of prediction (CV(pred) ) of 6.2%. When all samples within the whole concentration range are included, the SEP increases to 20.2 mg/dL, and hence the CV(pred) to 10.6% due to a nonlinear signal dependence on glucose concentration. A detection limit for glucose of 0.7 ng with a signal-to-noise ratio of 10 was obtained.


Asunto(s)
Glucosa/análisis , Espectroscopía Infrarroja por Transformada de Fourier/métodos , Reproducibilidad de los Resultados , Espectroscopía Infrarroja por Transformada de Fourier/instrumentación , Agua