Your browser doesn't support javascript.
loading
Mostrar: 20 | 50 | 100
Resultados 1 - 1 de 1
Filtrar
Más filtros

Bases de datos
Idioma
Tipo del documento
Asunto de la revista
País de afiliación
Intervalo de año de publicación
1.
Hell J Nucl Med ; 23(1): 97-107, 2020.
Artículo en El | MEDLINE | ID: mdl-32361720

RESUMEN

In this study, the properties of the reconstruction algorithm Iterative in comparison with the FORE-Iterative using 18F-FDG PET/CT data were evaluated. The study was conducted in the Department of Nuclear Medicine of Evangelismos Athens General Hospital, in which data from 9 patients were collected and reconstructed with both algorithms, Iterative and FORE-Iterative. For each patient, the image quality was assessed using parameters such as SUVmax, SUVpeak, SUVmean, signal-to-noise ratio (SNR) and contrast-to-noise ratio (CNR). Regions of interest were defined based on medical expertise focusing on the tumor areas. Based on the results of this study, images reconstructed with Iterative algorithm have better image quality compared to the images reconstructed with FORE-Iterative algorithm, considering that this particular method improves SNR and CNR. Using the Iterative algorithm, the image becomes sharper due to reduced noise resulting in safer clinical observations, and as a result a more accurate diagnosis.


Asunto(s)
Algoritmos , Procesamiento de Imagen Asistido por Computador/métodos , Tomografía Computarizada por Tomografía de Emisión de Positrones , Humanos , Relación Señal-Ruido