Your browser doesn't support javascript.
loading
Mostrar: 20 | 50 | 100
Resultados 1 - 2 de 2
Filtrar
Más filtros

Bases de datos
Tipo del documento
País de afiliación
Intervalo de año de publicación
1.
Cardiovasc Revasc Med ; 10(1): 45-8, 2009.
Artículo en Inglés | MEDLINE | ID: mdl-19159854

RESUMEN

BACKGROUND: Mechanical trauma caused by PCI is a primary reason for restenosis and subsequent target lesion revascularization (TLR). To minimize this trauma, we developed a computerized angioplasty pressure sensor and inflator device (CAPSID) for gradual inflation. The objective of this prospective randomized study was to examine whether use of CAPSID reduces early and late cardiac events in patients undergoing PCI. METHODS: Patients undergoing PCI were eligible and randomized to CAPSID or standard balloon inflation (plain old balloon angioplasty). In the CAPSID group, a slow, gradual balloon inflation was performed by a personal computer. Stenting was used in both groups only for suboptimal results. Patients with total occlusions and vein grafts were excluded. Clinical follow-up for major adverse cardiac events (MACE) was performed at 6 and 12 months, with repeat coronary angiography performed for clinical symptoms or positive stress testing. RESULTS: A total of 234 patients completed the study. At 1-year follow-up, the CAPSID group had a significantly lower rate of MACE (21% vs. 37%, P<.005). In patients who underwent angiography, there was a significantly lower rate of restenosis in the CAPSID group (20.2% vs. 35.5%). The reduction in TLR was even more pronounced in the subgroup undergoing stenting (8% vs. 24%; P<.001). CONCLUSIONS: We conclude that gradual computerized balloon inflation is more effective than standard manual balloon inflation in reducing adverse coronary events. The combination of CAPSID and subsequent stent deployment was especially effective in reducing TLR.


Asunto(s)
Angioplastia Coronaria con Balón/métodos , Enfermedades Cardiovasculares/prevención & control , Estenosis Coronaria/terapia , Terapia Asistida por Computador , Anciano , Angioplastia Coronaria con Balón/efectos adversos , Angioplastia Coronaria con Balón/instrumentación , Enfermedades Cardiovasculares/diagnóstico por imagen , Enfermedades Cardiovasculares/etiología , Angiografía Coronaria , Reestenosis Coronaria/etiología , Reestenosis Coronaria/prevención & control , Estenosis Coronaria/diagnóstico por imagen , Diseño de Equipo , Femenino , Humanos , Masculino , Persona de Mediana Edad , Estudios Prospectivos , Stents , Factores de Tiempo , Transductores de Presión , Resultado del Tratamiento
2.
Am J Cardiol ; 104(2): 223-6, 2009 Jul 15.
Artículo en Inglés | MEDLINE | ID: mdl-19576351

RESUMEN

Mechanical trauma caused by percutaneous coronary intervention is a major factor contributing to subsequent cardiac events, restenosis, and the need for target lesion revascularization (TLR). To minimize this trauma, we developed a Computerized Angioplasty Pressure Sensor and Inflator Device (CAPSID) for gradual inflation. The objective of the present prospective randomized study was to examine whether the use of this novel device reduced TLR, as well as cardiac events, in patients undergoing stenting. Patients undergoing coronary stenting were eligible and randomized to receive CAPSID or standard manual percutaneous coronary intervention. In the CAPSID group, slow, gradual balloon inflation was performed using a personal computer. Patients with acute ST-elevation myocardial infarction or the need for percutaneous coronary intervention for total occlusions, left main disease, and vein grafts were excluded. Clinical follow-up for major adverse cardiac events, including death, acute myocardial infarction, and TLR, was performed at 12 months. A total of 310 patients were enrolled in the study. No significant differences were found in the clinical characteristics between the CAPSID and control groups. At 1 year of follow-up, the CAPSID group had had a significantly lower rate of major adverse cardiac events (8% vs 18%, p <0.01) driven by significantly lower rates of acute myocardial infarction (1% vs 7%, p <0.01) and TLR (5% vs 12%, p <0.05). In conclusion, gradual computerized balloon inflation using CAPSID as a platform for angioplasty and stenting significantly reduced TLR and major adverse cardiac events at 1 year in patients undergoing coronary stenting. The use of this novel device may improve outcomes in patients undergoing coronary stenting.


Asunto(s)
Angioplastia Coronaria con Balón/instrumentación , Enfermedad de la Arteria Coronaria/terapia , Infarto del Miocardio/terapia , Stents , Terapia Asistida por Computador/instrumentación , Angioplastia de Balón/instrumentación , Enfermedad de la Arteria Coronaria/mortalidad , Femenino , Humanos , Masculino , Persona de Mediana Edad , Análisis Multivariante , Infarto del Miocardio/mortalidad , Estudios Prospectivos , Resultado del Tratamiento
SELECCIÓN DE REFERENCIAS
DETALLE DE LA BÚSQUEDA