1.
Pacing Clin Electrophysiol ; 40(2): 108-114, 2017 Feb.
Artículo en Inglés | MEDLINE | ID: mdl-27990645

RESUMEN

BACKGROUND: Automated external defibrillators can provide life-saving therapies to treat ventricular fibrillation. We developed a prototype unit that can deliver a unique shock waveform produced by four independent capacitors that is delivered through two shock vectors, with the rationale of providing more robust shock pathways during emergent defibrillation. We describe the initial testing and feasibility of this unique defibrillation unit, features of which may enable downsizing of current defibrillator devices. METHODS: We tested our defibrillation unit in four large animal models (two canine and two swine) under general anesthesia. Experimental defibrillation thresholds (DFT) were obtained by delivery of a unique waveform shock pulse via a dual-vector pathway with four defibrillation pads (placed across the chest). DFTs were measured and compared with those of a commercially available biphasic defibrillator (Zoll M series, Zoll Medical, Chelmsford, MA, USA) tested in two different vectors. Shocks were delivered after 10 seconds of stable ventricular fibrillation and the output characteristics and shock outcome recorded. Each defibrillation series used a step-down to failure protocol to define the defibrillation threshold. RESULTS: A total of 96 shocks were delivered during ventricular fibrillation in four large animals. In comparison to the Zoll M series, which delivered a single-vector, biphasic shock, the energy required for successful defibrillation using the unique dual-vector biphasic waveform did not differ significantly (P = 0.65). CONCLUSIONS: Our early findings support the feasibility of a unique external defibrillation unit using a dual-vector biphasic waveform approach. This warrants further study to leverage this unique concept and work toward a miniaturized, portable shock delivery system.


Asunto(s)
Desfibriladores , Electrodos , Terapia Asistida por Computador/instrumentación , Fibrilación Ventricular/prevención & control , Animales , Perros , Diseño de Equipo , Análisis de Falla de Equipo , Estudios de Factibilidad , Humanos , Miniaturización , Proyectos Piloto , Porcinos , Evaluación de la Tecnología Biomédica , Terapia Asistida por Computador/métodos , Resultado del Tratamiento
SELECCIÓN DE REFERENCIAS
DETALLE DE LA BÚSQUEDA