Guang Pu Xue Yu Guang Pu Fen Xi ; 29(5): 1431-3, 2009 May.
Artículo en Zh | MEDLINE | ID: mdl-19650507

RESUMEN

In the present study, the detection limit of atomic fluorescence spectrometry (AFS-930) was decreased to 2 ng x L(-1) (n=6) based on several optimizing modifications, including that the sub-high voltage of photomultiplier tube and the current of hollow-cathode lamp were elevated to 280 V and to 30 mA, respectively, and the height of atomization cell was set as 10 mm; In addition, the concentration of KBH4 was decreased to 0.5% (KOH 0.2%). With the optimized parameters, a good standard curve of Hg concentration versus intensity of fluorescence (If) could been obtained readily, after that, a 4-ng x L(-1)-Hg water samples was measured accurately with a little relative standard deviation (RSD) of <5%, while for approximately 2-ng x L(-1)-Hg waters the RSD varied within a wide range of 10.9%-27.2%, likely due to the absorption of Hg by polyethylene vessels used in this study and/or due to the contamination by analysis grade reagents used in this study. By using low-absorption polytetrafluoroethylene (PTFE) materials and the guaranteed reagents, the instrument detection limit was further decreased to 1 ng x L(-1) (n=10).


Asunto(s)
Límite de Detección , Espectrometría de Fluorescencia/instrumentación , Borohidruros/química , Calibración , Mercurio/análisis , Polietileno/química , Agua/química