1.
Eur J Cardiothorac Surg ; 30(1): 196-8, 2006 Jul.
Artículo en Inglés | MEDLINE | ID: mdl-16730192

RESUMEN

The following case report describes an unusual example of aortic valve damage caused by iatrogenic stent-graft rupture (disconnection of the proximal uncovered part of a stent-graft) during delivery of a proximal extension, resulting in the displacement of a stent-graft wire to the aortic root. The wire was extracted under cardiopulmonary bypass, using circulatory arrest, and the damaged aortic valve replaced by a mechanical valve.


Asunto(s)
Insuficiencia de la Válvula Aórtica/etiología , Prótesis Vascular/efectos adversos , Stents/efectos adversos , Adulto , Aneurisma Falso/diagnóstico por imagen , Aneurisma Falso/etiología , Aneurisma de la Aorta Torácica/diagnóstico por imagen , Aneurisma de la Aorta Torácica/cirugía , Válvula Aórtica/lesiones , Falla de Equipo , Implantación de Prótesis de Válvulas Cardíacas , Humanos , Masculino , Tomografía Computarizada por Rayos X