1.
J Cardiothorac Vasc Anesth ; 37(3): 392-398, 2023 03.
Artículo en Inglés | MEDLINE | ID: mdl-36577633

RESUMEN

OBJECTIVE: The aim of this study was to assess the reproducibility of the new viscoelastic analyzer ClotPro, and compare the parameters this system produces with the ROTEM delta system in patients undergoing cardiac surgery with cardiopulmonary bypass (CPB). DESIGN: A prospective and observational study. SETTING: At a university hospital. PARTICIPANTS: Forty adult patients undergoing cardiac surgery with CPB. INTERVENTIONS: Correlations were calculated between ClotPro and ROTEM delta parameters. MEASUREMENTS AND MAIN RESULTS: The ClotPro showed a high reproducibility in most of the parameters of each test; whereas ROTEM delta, although showing a low coefficient of variation in the parameters related to clot firmness, showed a high variability in the coagulation times. Excellent correlations were observed in most of the parameters of each test between ROTEM delta and ClotPro (≥0.93). However, a moderate correlation was obtained between the clotting time of the EXTEM and the EX-test (0.54). The concordance of amplitudes at different times within each test was almost 100% on both thromboelastometers. Regarding absolute differences in the test results, most of the measurements showed significant differences (p < 0.0001) between both devices. CONCLUSIONS: ClotPro can be used as an alternative to ROTEM delta to evaluate coagulation function in cardiac surgery, but specific reference ranges need to be established first.


Asunto(s)
Procedimientos Quirúrgicos Cardíacos , Tromboelastografía , Adulto , Humanos , Tromboelastografía/métodos , Estudios Prospectivos , Puente Cardiopulmonar/métodos , Reproducibilidad de los Resultados
2.
Clin Chem Lab Med ; 57(6): 883-890, 2019 05 27.
Artículo en Inglés | MEDLINE | ID: mdl-30530897

RESUMEN

Background Optimal haemostasis management in orthotropic liver transplant (OLT) could reduce blood loss and transfusion volume, improve patient outcomes and reduce cost. Methods We performed a study including 336 OLTs to evaluate the clinical and cost effectiveness of a new point-of-care (POC)-based haemostatic management approach in OLT patients. Results In terms of health benefit we found that the new approach showed a significant reduction in transfusion requirements (red blood cell transfusion units were reduced from 5.3±4.6 to 2.8±2.9 [p<0.001], free frozen plasma from 3.1±3.3 to 0.4±1.0 [p<0.001] and platelets from 2.9±3.9 to 0.4±0.9 [p<0.001], transfusion avoidance, 9.7% vs. 29.1% [p<0.001] and massive transfusion, 14.5% vs. 3.8% [p=0.001]); we also found a significant improvement in patient outcomes, such, reoperation for bleeding or acute-kidney-failure (8.3% vs. 2.4%, p=0.015; 33.6% vs. 5.4%, p<0.001), with a significant reduction in the length of the hospital total stay (40.6±13.8 days vs. 38.2±14.4 days, p=0.001). The lowest cost incurred was observed with the new approach (€73,038.80 vs. €158,912.90) with significant patient saving associated to transfusion avoidance (€1278.36), ICU-stay (€3037.26), total-stay (€3800.76) and reoperation for bleeding (€80,899.64). Conclusions POC haemostatic monitoring during OLT is cost effective.


Asunto(s)
Análisis Costo-Beneficio , Enfermedad Hepática en Estado Terminal/terapia , Trasplante de Hígado , Sistemas de Atención de Punto/economía , Pruebas de Coagulación Sanguínea , Transfusión de Eritrocitos , Humanos , Tiempo de Internación
4.
Clin Chim Acta ; 446: 277-83, 2015 Jun 15.
Artículo en Inglés | MEDLINE | ID: mdl-25916692

RESUMEN

BACKGROUND: Optimal haemostasis management can improve patient outcomes and reduce blood loss and transfusion volume in orthotopic-liver-transplant (OLT). METHODS: We performed a prospective study including 200 consecutive OLTs. The first 100 patients were treated according to the clinic's standards and the next 100 patients were treated using the new point-of-care (POC)-based haemostasis management strategy. Transfusion parameters and other outcomes were compared between groups. RESULTS: Transfusion requirements were reduced in the POC group. The median and IQR of red-blood-cells (RBC) transfusion units were reduced from 5 [2-8] to 3 [0-5] (p < 0.001), plasma from 2 [0-4] to 0 (p < 0.001), and platelets from 1 [0-4] to 0 [0-1] (p < 0.001), into the POC group only four patients received tranexamic acid and fibrinogen transfusion rate was 1.13 ± 1.44 g (p = 0.001). We also improved the incidence of transfusion avoidance, 5% vs. 24% (p < 0.001) and reduced the incidence of massive transfusion (defined as the transfusion of more than 10 RBC units), 13% vs. 2% (p = 0.005). We also observed a relationship between RBC transfusion requirements and preoperative haemoglobin, and between platelet transfusion and preoperative fibrinogen levels. The incidence of postoperative complications, such as, reoperation for bleeding, acute-kidney-failure or haemodynamic instability was significantly lower (13.0% vs. 5%, p = 0.048, 17% vs. 2%, p < 0.001, and 29% vs. 16%, p = 0.028). Overall, blood product transfusion was associated with increased risk of postoperative complications. CONCLUSIONS: A haemostatic therapy algorithm based on POC monitoring reduced transfusion and improved outcome in OLT.


Asunto(s)
Lesión Renal Aguda/prevención & control , Hemorragia/prevención & control , Técnicas Hemostáticas , Trasplante de Hígado/efectos adversos , Complicaciones Posoperatorias , Lesión Renal Aguda/etiología , Lesión Renal Aguda/patología , Transfusión de Eritrocitos/economía , Transfusión de Eritrocitos/estadística & datos numéricos , Femenino , Fibrinógeno/metabolismo , Hemoglobinas/metabolismo , Hemorragia/etiología , Hemorragia/patología , Hemorragia/terapia , Hemostasis , Humanos , Masculino , Persona de Mediana Edad , Transfusión de Plaquetas/economía , Transfusión de Plaquetas/estadística & datos numéricos , Sistemas de Atención de Punto , Estudios Prospectivos , Riesgo , Resultado del Tratamiento