1.
J Heart Valve Dis ; 24(2): 214-6, 2015 Mar.
Artículo en Inglés | MEDLINE | ID: mdl-26204688

RESUMEN

A case is presented of the early degeneration of a 21 mm SOLO stentless valve concomitant with severe mitral regurgitation (MR). Transcatheter valve-in-valve implantation was considered in this high-risk case (logistic EuroSCORE 29.3%), but was dismissed because of the risk of coronary occlusion, an absence of visual landmarks, and the impossibility to treat the MR. Following the implantation of a 27 mm Medtronic Hancock II mitral bioprosthesis, the leaflets of the SOLO valve were removed, and a 19 mm 3f Enable sutureless bioprosthesis delivered into the remaining sewing belt of the stentless valve. The total cross-clamp time was 64 min. No aortic paravalvular leakage was detected at discharge and early follow up (four months); the mean and peak transvalvular aortic gradients were 13 and 23 mmHg, respectively, and the left ventricular ejection fraction 60%. A sutureless strategy simplified the management of this high-risk case.


Asunto(s)
Estenosis de la Válvula Aórtica/cirugía , Bioprótesis , Implantación de Prótesis de Válvulas Cardíacas/métodos , Prótesis Valvulares Cardíacas , Anciano , Estenosis de la Válvula Aórtica/fisiopatología , Femenino , Humanos , Diseño de Prótesis