1.
Scand J Clin Lab Invest ; 79(7): 468-474, 2019 Nov.
Artículo en Inglés | MEDLINE | ID: mdl-31460810

RESUMEN

The purpose of this study was to evaluate the analytical performances of Sysmex UF-5000 and Dirui FUS-200 and to compare the results with manual microscopy and between each other. Two hundred fifty urine samples were analyzed for evaluation. Mid-stream specimens were studied sequentially using Dirui FUS-200 and Sysmex UF-5000, and also with manual microscopy within one hour. The physical and chemical components of urinalysis, and sediment results were investigated. The precision results of the FUS-200 and UF-5000 for WBCs, RBCs, and ECs were acceptable. The both analyzers demonstrated good linearity (r > 0.97), with no carry-over. The comparisons of FUS-200 and UF-5000 with manual microscopy for RBCs, WBCs, and ECs on 250 samples exhibited good agreement with little bias (R > 0.780). Only, the moderate agreements were obtained for calcium oxalate for both analyzers (R = 0.512, and 0.648, respectively). The sensitivities of the FUS-200 and UF-5000 were 75.8% and 86.8%, with specificities of 92.3% and 87.8% for WBCs, for RBCs the sensitivities were 91.1%, and 84.4% with specificities of 82.2%, and 89.6% for both analyzers. Kappa values of the UF-5000 were higher than FUS-200 for WBCs, RBCs, ECs, and calcium oxalate. The FUS-200 and UF-5000 urine analyzers, are both accurate, very precise systems and can be safely used in clinical laboratories. However, due to the technological characteristics of the UF-5000 analyzer, its positive impacts on the morphologic recognition and enumeration of RBCs and WBCs should be taken into account, particularly in university hospital laboratories with high patient volumes.


Asunto(s)
Citometría de Flujo/instrumentación , Urinálisis/instrumentación , Orina/citología , Automatización de Laboratorios , Eritrocitos , Citometría de Flujo/métodos , Humanos , Procesamiento de Imagen Asistido por Computador , Leucocitos , Microscopía , Sensibilidad y Especificidad , Urinálisis/métodos
2.
Scand J Clin Lab Invest ; 74(5): 414-7, 2014 Aug.
Artículo en Inglés | MEDLINE | ID: mdl-24693995

RESUMEN

OBJECTIVES: The microscopic analysis of urine is essential for the diagnosis of patients with urinary tract infections. Quantitative urine culture is the 'gold standard' method for definitive diagnosis of urinary-tract infections, but it is labor-intensive, time consuming, and does not provide the same-day results. The aim of this study was to evaluate the analytical and diagnostic performance of the FUS200 (Changchun Dirui Industry, China), a new urine sedimentation analyzer in comparison to urine culture as the reference method. METHODS: We evaluated 1000 urine samples, submitted for culture and urine analysis with a preliminary diagnosis of urinary-tract infection. Cut-off values for the FUS200 were determined by comparing the results with urine cultures. The cut-off values by the receiver operating characteristic (ROC) curve technique, sensitivity, and specificity were calculated for bacteria and white blood cells (WBCs). RESULTS: Among the 1000 urine specimens submitted for culture, 637 cultures (63.7%) were negative, and 363 were (36.3%) positive. The best cut-off values obtained from ROC analysis were 16/µL for bacteriuria (sensitivity: 82.3%, specificity: 58%), and 34/µL for WBCs (sensitivity: 72.3%, specificity: 65.2%). The area under the curve (AUC) for the bacteria and WBCs count were 0.71 (95% CI: 0.67-0.74) and, 0.72 (95% CI: 0.69-0.76) respectively. CONCLUSIONS: The most important requirement of a rapid diagnostic screening test is sensitivity, and, in this perspective, an unsatisfactory sensitivity by using bacteria recognition and quantification performed by the FUS200 analyzer has been observed. After further technical improvements in particle recognition and laboratory personnel training, the FUS200 might show better results.


Asunto(s)
Bacteriuria/orina , Urinálisis/instrumentación , Adolescente , Adulto , Anciano , Anciano de 80 o más Años , Área Bajo la Curva , Carga Bacteriana , Bacteriuria/inmunología , Bacteriuria/microbiología , Niño , Preescolar , Femenino , Humanos , Lactante , Recién Nacido , Recuento de Leucocitos , Leucocitos/inmunología , Masculino , Persona de Mediana Edad , Curva ROC , Valores de Referencia , Estudios Retrospectivos , Adulto Joven
3.
Medeni Med J ; 34(3): 244-251, 2019.
Artículo en Inglés | MEDLINE | ID: mdl-32821445

RESUMEN

OBJECTIVE: Microscopic examination of urine sediment is necessary for evaluation of renal and urinary tract diseases. In this study, we evaluated and compared analytic and diagnostic performances of DIRUI FUS-200 and a new image-based automated urine sediment analyzer Urised 3. METHOD: A total of 440 urine samples, submitted to our laboratory, were evaluated by two automated urine sediment analyzers and a standardized manual microscopy. Precision, linearity and method comparison studies were performed according to CLSI guidelines. RESULTS: Considering the red blood cell (RBC) and white blood cell (WBC) counts, strong correlations existed between FUS-200 and manual microscopy (r=0.993 vs 0.861), Urised 3 and manual microscopy (r=0.962 vs 0.818), FUS200 and Urised 3 (r=0.961 vs 0.961). Clinical non-concordance ranged from 7% to 14.16% among all methods. CONCLUSIONS: The concordance between the analyzers and manual microscopy for WBC was better than that of RBC. The concordance between the two analyzers was better for WBC and RBC, with respect to the manual microscopy. Although the Urised 3, FUS-200 and manual microscopy counts were in agreement; confirmation of the results of automated analyzers with manual microscopy is particularly helpful, for pathological samples with near cut-off values.

4.
Front Med (Lausanne) ; 5: 353, 2018.
Artículo en Inglés | MEDLINE | ID: mdl-30619863

RESUMEN

Urinary tract infection is the most common human infection with a high morbidity. In primary care and hospital services, conventional urine culture is a key part of infection diagnosis but results take at least 24 h. Therefore, a rapid and reliable screening method is still needed to discard negative samples as quickly as possible and to reduce the laboratory workload. In this aspect, this study aims to compare the diagnostic performance between Sysmex UF-1000i and FUS200 systems in comparison to urine culture as the gold standard. From March to June 2016, 1,220 urine samples collected at the clinical microbiology laboratory of the "Miguel Servet" hospital were studied in parallel with both analysers, and some technical features were evaluated to select the ideal equipment. The most balanced cut-off values taking into account bacteria or leukocyte counts were 138 bacteria/µL or 119.8 leukocyte/µL for the UF-1000i (95.3% SE and 70.4% SP), and 5.7 bacteria/µL or 4.3 leukocyte/µL for the FUS200 (95.8% SE and 44.4% SP). The reduction of cultured plates was 37.4% with the FUS200 and 58.3% with the UF-1000i. This study shows that both techniques improve the workflow in the laboratory, but the UF-1000i has the highest specificity at any sensitivity and the FUS200 needs a shorter processing time.