Your browser doesn't support javascript.
loading
Mostrar: 20 | 50 | 100
Resultados 1 - 1 de 1
Filtrar
Más filtros

Bases de datos
Tipo del documento
Asunto de la revista
País de afiliación
Intervalo de año de publicación
1.
Eur Spine J ; 31(11): 3098-3108, 2022 11.
Artículo en Inglés | MEDLINE | ID: mdl-36149493

RESUMEN

STUDY DESIGN: Prospective multi-center study. OBJECTIVE: The study aimed to evaluate the accuracy of pedicle screw placement using a skin marker-based optical surgical navigation system for minimal invasive thoraco-lumbar-sacral pedicle screw placement. METHODS: The study was performed in a hybrid Operating Room with a video camera-based navigation system integrated in the imaging hardware. The patient was tracked with non-invasive skin markers while the instrument tracking was via an on-shaft optical marker pattern. The screw placement accuracy assessment was performed by three independent reviewers, using the Gertzbein grading. The screw placement time as well as the staff and patient radiation doses was also measured. RESULTS: In total, 211 screws in 39 patients were analyzed for screw placement accuracy. Of these 32.7% were in the thoracic region, 59.7% were in the lumbar region, and 7.6% were in the sacral region. An overall accuracy of 98.1% was achieved. No screws were deemed severely misplaced (Gertzbein grading 3). The average time for screw placement was 6 min and 25 secs (± 3 min 33 secs). The average operator radiation dose per subject was 40.3 µSv. The mean patient effective dose (ED) was 11.94 mSv. CONCLUSION: Skin marker-based ON can be used to achieve very accurate thoracolumbarsacral pedicle screw placements.


Asunto(s)
Tornillos Pediculares , Fusión Vertebral , Cirugía Asistida por Computador , Humanos , Estudios Prospectivos , Región Sacrococcígea , Vértebras Lumbares/diagnóstico por imagen , Vértebras Lumbares/cirugía , Imagenología Tridimensional/métodos , Cirugía Asistida por Computador/métodos , Fusión Vertebral/métodos
SELECCIÓN DE REFERENCIAS
DETALLE DE LA BÚSQUEDA