Your browser doesn't support javascript.
loading
Integration of a contactless conductivity detector into a commercial capillary cassette. Detection of inorganic cations and catecholamines.
Vuorinen, Pasi S; Jussila, Matti; Sirén, Heli; Palonen, Sami; Riekkola, Marja-Liisa.
Afiliación
  • Vuorinen PS; Laboratory of Analytical Chemistry, Department of Chemistry, University of Helsinki, P.O. Box 55, FIN-00014 University of Helsinki, Finland.
J Chromatogr A ; 990(1-2): 45-52, 2003 Mar 21.
Article en En | MEDLINE | ID: mdl-12685582
ABSTRACT
A contactless conductivity detector integrated into the capillary cassette of Agilent (3D)CE equipment is described. The detector is user-friendly, compact and easily modified. The UV detector of the (3D)CE equipment is available parallel with the contactless conductivity detector increasing the detection power. Two electrolyte solutions, 2-(N-morpholino)ethanesulfonic acid-histidine solution (20 mM, pH 6.0) and ammonium acetate (10 mM, pH 4.0), were used as the separation media for inorganic cations and organic catecholamines, respectively. The detection limit for all metal cations except barium was under 0.5 mg/l, and that for four catecholamines was ca. 10 mg/l. This last value was the same order of magnitude as achieved with parallel UV detection.
Asunto(s)
Buscar en Google
Bases de datos: MEDLINE Asunto principal: Catecolaminas / Cationes / Electroforesis Capilar / Conductividad Eléctrica Tipo de estudio: Diagnostic_studies Idioma: En Revista: J Chromatogr A Año: 2003 Tipo del documento: Article País de afiliación: Finlandia
Buscar en Google
Bases de datos: MEDLINE Asunto principal: Catecolaminas / Cationes / Electroforesis Capilar / Conductividad Eléctrica Tipo de estudio: Diagnostic_studies Idioma: En Revista: J Chromatogr A Año: 2003 Tipo del documento: Article País de afiliación: Finlandia