Comparison of two automated nephelometers.
Emancipator, K; Elin, R J; Fleisher, T A.
Afiliación
  • Emancipator K; Clinical Pathology Department, Warren G. Magnuson Clinical Center, National Institutes of Health, Bethesda, Maryland.
J Clin Lab Anal ; 6(6): 399-404, 1992.
Article en En | MEDLINE | ID: mdl-1432366
ABSTRACT
The intercalibration precision and linearity were determined for two representative analytes, apolipoprotein A1 (apo A1) and immunoglobulin G (IgG), on the Beckman Array and the Behring Nephelometer 100 (BN-100). For two of nine samples analyzed, poor precision was observed for IgG with the Array. The poor precision for these two samples is attributed to large systematic shifts. A statistical non-linearity was observed for apo A1 with the BN-100, but this non-linearity does not exclude use of this assay for clinical diagnosis. Method comparisons were done for 12 analytes apo A1, apo B, IgG, IgA, IgM, IgE, C3, C4, albumin, C-reactive protein, alpha-1-antitrypsin, and ceruloplasmin. These comparisons showed proportional biases of > 10% for seven of 12 analytes. Furthermore, correlation coefficients were < 0.96 for seven of 12 analytes. We conclude that comparing results obtained from the two different nephelometers is of only limited value for the individual patient.
Asunto(s)
Buscar en Google
Bases de datos: MEDLINE Asunto principal: Proteínas / Nefelometría y Turbidimetría Tipo de estudio: Diagnostic_studies / Evaluation_studies Límite: Humans Idioma: En Revista: J Clin Lab Anal Asunto de la revista: TECNICAS E PROCEDIMENTOS DE LABORATORIO Año: 1992 Tipo del documento: Article
Buscar en Google
Bases de datos: MEDLINE Asunto principal: Proteínas / Nefelometría y Turbidimetría Tipo de estudio: Diagnostic_studies / Evaluation_studies Límite: Humans Idioma: En Revista: J Clin Lab Anal Asunto de la revista: TECNICAS E PROCEDIMENTOS DE LABORATORIO Año: 1992 Tipo del documento: Article