Monte carlo simulations of dose from microCT imaging procedures in a realistic mouse phantom.
Taschereau, Richard; Chow, Patrick L; Chatziioannou, Arion F.
Afiliación
  • Taschereau R; The Crump Institute for Molecular Imaging, Department of Molecular and Medical Pharmacology, University of California School of Medicine, 700 Westwood Boulevard, Los Angeles, California 90095, USA. rtaschereau@mednet.ucla.edu
Med Phys ; 33(1): 216-24, 2006 Jan.
Article en En | MEDLINE | ID: mdl-16485428
ABSTRACT
The purpose of this work was to calculate radiation dose and its organ distribution in a realistic mouse phantom from micro-computed tomography (microCT) imaging protocols. CT dose was calculated using GATE and a voxelized, realistic phantom. The x-ray photon energy spectra used in simulations were precalculated with GATE and validated against previously published data. The number of photons required per simulated experiments was determined by direct exposure measurements. Simulated experiments were performed for three types of beams and two types of mouse beds. Dose-volume histograms and dose percentiles were calculated for each organ. For a typical microCT screening examination with a reconstruction voxel size of 200 microm, the average whole body dose varied from 80 mGy (at 80 kVp) to 160 mGy (at 50 kVp), showing a strong dependence on beam hardness. The average dose to the bone marrow is close to the soft tissue average. However, due to dose nonuniformity and higher radiation sensitivity, 5% of the marrow would receive an effective dose about four times higher than the average. If CT is performed longitudinally, a significant radiation dose can be given. The total absorbed radiation dose is a function of milliamperes-second, beam hardness, and desired image quality (resolution, noise and contrast). To reduce dose, it would be advisable to use the hardest beam possible while maintaining an acceptable contrast in the image.
Asunto(s)

Texto completo: 1 Bases de datos: MEDLINE Asunto principal: Radiometría / Tomografía Computarizada por Rayos X / Imagen de Cuerpo Entero / Modelos Biológicos Tipo de estudio: Health_economic_evaluation / Risk_factors_studies Límite: Animals Idioma: En Revista: Med Phys Año: 2006 Tipo del documento: Article País de afiliación: Estados Unidos

Texto completo: 1 Bases de datos: MEDLINE Asunto principal: Radiometría / Tomografía Computarizada por Rayos X / Imagen de Cuerpo Entero / Modelos Biológicos Tipo de estudio: Health_economic_evaluation / Risk_factors_studies Límite: Animals Idioma: En Revista: Med Phys Año: 2006 Tipo del documento: Article País de afiliación: Estados Unidos