Dose distribution around a needle-like anode X-ray tube: dye-film vs. planar thermoluminescent detectors.
Budzanowski, M; Olko, P; Marczewska, B; Czopyk, L; Slapa, M; Stras, W; Traczyk, M; Talejko, M.
Afiliación
  • Budzanowski M; Institute of Nuclear Physics, Polish Academy of Sciences (IFJ PAN), ul. Radzikowskiego 152, PL-31-342 Kraków, Poland. maciej.budzanowski@ifj.edu.pl
Radiat Prot Dosimetry ; 120(1-4): 117-20, 2006.
Article en En | MEDLINE | ID: mdl-16614087
ABSTRACT
The dosimetry around the X-ray tube with a needle-like anode (NAXT), developed at the Institute of Nuclear Studies, for interstitial brachytherapy has been performed using (1) dye films (Gafchromic XR-T), (2) large-area thermoluminescent (TL) detectors--prepared either by gluing TL powder onto thin Al foil (so-called planar detectors with spatial resolution of 0.1 mm) and (3) miniature (2 mm diameter and 0.5 mm thick) TL detectors. The measurements were performed in following geometries. (1) Needle inside a PMMA cylinder--the planar TL detector mounted on the surface of the cylinder. (2) Needle inside a thick block of PMMA and TL detector mounted vertically 7 mm from needle axis. TL detectors were read with the planar (2D) thermoluminescence reader, developed at IFJ, with a sensitive CCD (charge couple device) camera. Gafchromic films were evaluated with a system based on Agfa Arcus 1200 scanner and calibrated with X rays (35 kV) filtered with 0.03 mm Mo and with Co-60 photons. The intensity distribution of TL light on the planar detector was calibrated in terms of absorbed dose to water, using (137)Cs gamma-rays. TL planar detectors seem to be a promising tool for 2D dosimetry of miniature X-ray sources. Obtained results for TLDs and Gafchromic films seem to be comparable but differences have been found. Both methods are useful for measurements of dose distribution around the NAXT X-rays source.
Asunto(s)
Buscar en Google
Bases de datos: MEDLINE Asunto principal: Protección Radiológica / Dosimetría Termoluminiscente / Braquiterapia / Dosimetría por Película Tipo de estudio: Diagnostic_studies / Evaluation_studies Idioma: En Revista: Radiat Prot Dosimetry Año: 2006 Tipo del documento: Article País de afiliación: Polonia
Buscar en Google
Bases de datos: MEDLINE Asunto principal: Protección Radiológica / Dosimetría Termoluminiscente / Braquiterapia / Dosimetría por Película Tipo de estudio: Diagnostic_studies / Evaluation_studies Idioma: En Revista: Radiat Prot Dosimetry Año: 2006 Tipo del documento: Article País de afiliación: Polonia