A multi-element solid-phase extraction method for trace metals determination in environmental samples on Amberlite XAD-2000.
J Hazard Mater
; 146(1-2): 155-63, 2007 Jul 19.
Article
en En
| MEDLINE
| ID: mdl-17222968
A method for the preconcentration of some transition elements at trace level was proposed using a column filled with Amberlite XAD-2000 resin. Metal ions were adsorbed on XAD-2000 as their diethyldithiocarbamate chelates, then analytes retained on the resin were eluted by 1 mol L(-1) nitric acid in acetone and determined by flame atomic absorption spectrometry (FAAS). The influences of some analytical parameters including pH of sample solution, ligand amount, the type, concentration and volume of elution solution, flow rates of the sample and eluent solutions, adsorption capacity of the resin and sample volume on the preconcentration efficiency have been investigated. The influences of some matrix elements were also examined. The detection limit (N=20, 3 sigma) for Mn(II), Fe(II), Co(II), Cu(II), Cd(II), Zn(II), Pb(II) and Ni(II) were found as 0.20, 0.35, 0.25, 0.20, 0.20, 0.15, 0.45 and 0.25 microg L(-1), respectively. The validation of the procedure was carried out by analysis of certified reference materials. The proposed method was applied to natural waters and kale vegetable (Brassica oleracea var. acephala).
Buscar en Google
Bases de datos:
MEDLINE
Asunto principal:
Contaminantes Químicos del Agua
/
Brassica
/
Metales Pesados
País/Región como asunto:
Asia
Idioma:
En
Revista:
J Hazard Mater
Asunto de la revista:
SAUDE AMBIENTAL
Año:
2007
Tipo del documento:
Article
País de afiliación:
Turquía