High-resolution coronary imaging by optical coherence tomography: Feasibility, pitfalls and artefact analysis.
Motreff, Pascal; Levesque, Sébastien; Souteyrand, Géraud; Sarry, Laurent; Ouchchane, Lemlih; Citron, Bernard; Cassagnes, Jean; Lusson, Jean-René.
Afiliación
  • Motreff P; Department of Cardiology, Gabriel-Montpied Hospital, CHU Clermont-Ferrand, 63003 Clermont-Ferrand, France. pmotreff@chu-clermontferrand.fr
Arch Cardiovasc Dis ; 103(4): 215-26, 2010 Apr.
Article en En | MEDLINE | ID: mdl-20656632
ABSTRACT

BACKGROUND:

Optical coherence tomography is an imaging method that enables cardiologists to study atheromatous plaques, and to check the implantation and evolution of coronary stents. It is an invasive technique, providing high-resolution (10 microm) in vivo images, but with limitations and artefacts that need to be understood before the field of application can be extended.

AIM:

To determine the feasibility and limitations of optical coherence tomography coronary imaging from a single-centre experience.

METHODS:

We analysed the first 301 optical coherence tomography (version M2, LightLab Imaging) sequences obtained in our department from examination of 73 patients.

RESULTS:

Results showed that 92% of sequences for selected lesions were usable, with a mean examination time of 17 min. Only one complication occurred (ventricular fibrillation, reduced by external electroshock). In our registry, sequence quality depended on operator experience (improving after 20 examinations), and was impaired by artefacts, especially in right coronary analysis and in arteries of greater than 3.5 mm calibre.

CONCLUSIONS:

Proximal coronary occlusion and the distal flush quality currently required for quality imaging should no longer be indispensable with the new generation of optical coherence tomography systems.
Asunto(s)

Texto completo: 1 Bases de datos: MEDLINE Asunto principal: Enfermedad de la Arteria Coronaria / Artefactos / Vasos Coronarios / Tomografía de Coherencia Óptica Tipo de estudio: Prognostic_studies Límite: Adult / Aged / Female / Humans / Male / Middle aged País/Región como asunto: Europa Idioma: En Revista: Arch Cardiovasc Dis Asunto de la revista: ANGIOLOGIA / CARDIOLOGIA Año: 2010 Tipo del documento: Article País de afiliación: Francia

Texto completo: 1 Bases de datos: MEDLINE Asunto principal: Enfermedad de la Arteria Coronaria / Artefactos / Vasos Coronarios / Tomografía de Coherencia Óptica Tipo de estudio: Prognostic_studies Límite: Adult / Aged / Female / Humans / Male / Middle aged País/Región como asunto: Europa Idioma: En Revista: Arch Cardiovasc Dis Asunto de la revista: ANGIOLOGIA / CARDIOLOGIA Año: 2010 Tipo del documento: Article País de afiliación: Francia