Your browser doesn't support javascript.
loading
Image quality and radiation dose in 256-slice cardiac computed tomography: comparison of prospective versus retrospective image acquisition protocols.
Hosch, Waldemar; Heye, Tobias; Schulz, Felix; Lehrke, Stephanie; Schlieter, Martin; Giannitsis, Evangelos; Kauczor, Hans-Ulrich; Katus, Hugo A; Korosoglou, Grigorios.
Afiliación
  • Hosch W; University of Heidelberg, Department of Diagnostic and Interventional Radiology, Heidelberg, Germany.
Eur J Radiol ; 80(1): 127-35, 2011 Oct.
Article en En | MEDLINE | ID: mdl-20708867
ABSTRACT

PURPOSE:

To assess coronary artery image quality and patient radiation exposure in patients who underwent clinically indicated 256-slice CTA.

METHODS:

Consecutive patients (n=193) underwent 256-slice CTA, using (1) retrospective gating without radiation dose modulation, (2) retrospective gating with radiation dose modulation and (3) prospective gating. Image quality was determined by consensus of two experienced observers using a 5-grade scale. The effective dose was calculated.

RESULTS:

In all patients, CTA was performed without adverse events. Retrospective CTA was assessed in 39 patients with and 39 without dose modulation, while 115 patients underwent prospective CTA. Heart rate was related to image quality with all protocols (r=0.46, p<0.001). Up to a heart rate of 75 bpm no significant difference in overall image quality was observed for all three protocols, while no significant differences could be observed between retrospective CTA with and without dose modulation for any segments or heart rates. Prospective and retrospective CTA with dose modulation showed radiation savings of ∼75 % and ∼30 %, respectively compared to retrospective CTA without dose modulation (p<0.001).

CONCLUSIONS:

In patients with heart rates up to 75 bpm prospective CTA should be the first choice acquisition protocol. For heart rates >75 bpm, retrospective CTA with dose modulation should be considered.
Asunto(s)

Texto completo: 1 Bases de datos: MEDLINE Asunto principal: Dosis de Radiación / Enfermedad de la Arteria Coronaria / Angiografía Coronaria / Tomografía Computarizada Multidetector Tipo de estudio: Clinical_trials / Guideline Límite: Female / Humans / Male / Middle aged Idioma: En Revista: Eur J Radiol Año: 2011 Tipo del documento: Article País de afiliación: Alemania

Texto completo: 1 Bases de datos: MEDLINE Asunto principal: Dosis de Radiación / Enfermedad de la Arteria Coronaria / Angiografía Coronaria / Tomografía Computarizada Multidetector Tipo de estudio: Clinical_trials / Guideline Límite: Female / Humans / Male / Middle aged Idioma: En Revista: Eur J Radiol Año: 2011 Tipo del documento: Article País de afiliación: Alemania