Dual-source parallel RF transmission for clinical MR imaging of the spine at 3.0 T: intraindividual comparison with conventional single-source transmission.
Nelles, Michael; König, Roy S; Gieseke, Jürgen; Guerand-van Battum, Marjolijn M; Kukuk, Guido M; Schild, Hans H; Willinek, Winfried A.
Afiliación
  • Nelles M; Department of Radiology, University of Bonn Hospital and Medical School, Sigmund Freud Strasse 25, D-53105 Bonn, Germany. michael.nelles@ukb.uni-bonn.de
Radiology ; 257(3): 743-53, 2010 Dec.
Article en En | MEDLINE | ID: mdl-20858848
ABSTRACT

PURPOSE:

To prospectively and intraindividually compare single-source radiofrequency (RF) excitation and dual-source parallel RF excitation in 3.0-T magnetic resonance (MR) imaging of the spine. MATERIALS AND

METHODS:

Institutional review board approval and written informed patient consent were obtained. The RF power of a 3.0-T MR imaging system was distributed to two ports of the body coil of the system by using independent RF transmit channels. The maximum B(1) field strength for dual-source parallel RF excitation was maintained, as compared with single-source RF excitation. The repetition time was reduced according to the revised RF setup with dual-source parallel RF excitation while maintaining specific energy absorption limitations. Thirty patients were examined with and without dual-source parallel RF excitation. Diagnostic quality was assessed independently by two radiologists according to a four-point grading system. Image contrast ratios (CRs) were calculated between reference tissues and vertebrae for single-source RF excitation and dual-source parallel RF excitation.

RESULTS:

The mean acceleration achieved with dual-source parallel RF excitation was 36% (range, 18%-50%). The total imaging duration of a three-station total spinal examination was reduced by one-third by using dual-source parallel RF transmission. For all cases investigated, diagnostic image quality without significant differences between the two methods and with a good interobserver agreement was achieved (Kendall tau-b, 0.50-0.84). The observed image contrast changes were predominantly small (<0.10 in 15 of 24 CRs), though they were significantly different (P < .05).

CONCLUSION:

While shortening examination times by approximately one-third, the dual-source parallel RF transmission mode in MR imaging of the spine yielded diagnostic image quality comparable to that with the conventional single-source RF transmission mode.
Asunto(s)

Texto completo: 1 Bases de datos: MEDLINE Asunto principal: Enfermedades de la Columna Vertebral / Imagen por Resonancia Magnética / Aumento de la Imagen Tipo de estudio: Observational_studies Límite: Adult / Aged / Aged80 / Female / Humans / Male / Middle aged Idioma: En Revista: Radiology Año: 2010 Tipo del documento: Article País de afiliación: Alemania

Texto completo: 1 Bases de datos: MEDLINE Asunto principal: Enfermedades de la Columna Vertebral / Imagen por Resonancia Magnética / Aumento de la Imagen Tipo de estudio: Observational_studies Límite: Adult / Aged / Aged80 / Female / Humans / Male / Middle aged Idioma: En Revista: Radiology Año: 2010 Tipo del documento: Article País de afiliación: Alemania