Investigation of false-positive results given by the QuantiFERON-TB Gold In-Tube assay.
J Clin Microbiol
; 50(9): 3105-7, 2012 Sep.
Article
en En
| MEDLINE
| ID: mdl-22785197
ABSTRACT
We investigated a sudden increase in the rate of positive QuantiFERON-TB Gold In-Tube results from 10% to 31% at a U.S. academic institution. Direct comparison of the TB antigen tubes with tubes from a different lot number identified that a potential problem with the TB antigen vials in a certain tube lot was the likely cause of the elevated positive rate. The underlying defect remains unknown. This finding warrants refinement of quality control programs by the manufacturer and users.