Patent ductus arteriosus closure using the Amplatzer® vascular plug II for all anatomic variants.
Delaney, Jeffrey W; Fletcher, Scott E.
Afiliación
  • Delaney JW; Children's Hospital and Medical Center, Department of Pediatrics, Division of Pediatric Cardiology, University of Nebraska, Omaha, Nebraska 68114, USA. jdelaney@childrensomaha.org
Catheter Cardiovasc Interv ; 81(5): 820-4, 2013 Apr.
Article en En | MEDLINE | ID: mdl-23074167
ABSTRACT

OBJECTIVES:

To evaluate the safety and efficacy of the Amplatzer® Vascular Plug II (AVPII) for closure of the patent ductus arteriosus (PDA).

BACKGROUND:

The PDA has significant anatomic variation. No device has proven applicable to all PDAs. Previous case reports and small series have documented limited use of the AVPII for some PDA types. We describe the largest and most diverse experience using the AVPII.

METHODS:

A retrospective analysis of patients undergoing percutaneous PDA closure between 01/01/2009 and 05/01/2012 was performed. The PDA was characterized, measured, and the device chosen was listed. Deployment technique, complications and procedural results were recorded.

RESULTS:

Sixty-seven procedures were performed. The AVPII was utilized for 43 (64.2%), 15 (20.9%) had coils, 7 (10.4%) had the AGA duct occluder, and 3 (4.5%) were referred for surgery. The AVPII was placed in infants as young as 2 months and 4.2 kg. AVPII size ranged from 4 to 10 mm. All PDA types were closed. Retrograde and antegrade deployments were performed, using the outer disc as a "retention skirt" to secure the device and improve occlusion. Three patients were up-sized prior to release. All deployments were successful; 89% "in-lab" and 100% closure on postprocedural echocardiogram. There were no complications.

CONCLUSIONS:

We report the largest experience with the AVPII for PDA closure. The device was used in all morphologic types and small patients. It is low profile, easily repositioned, and had excellent results without complications. We contend that this is the most versatile device currently available.
Asunto(s)

Texto completo: 1 Bases de datos: MEDLINE Asunto principal: Cateterismo Cardíaco / Conducto Arterioso Permeable / Dispositivo Oclusor Septal Tipo de estudio: Diagnostic_studies / Observational_studies Límite: Humans / Infant Idioma: En Revista: Catheter Cardiovasc Interv Asunto de la revista: CARDIOLOGIA Año: 2013 Tipo del documento: Article País de afiliación: Estados Unidos

Texto completo: 1 Bases de datos: MEDLINE Asunto principal: Cateterismo Cardíaco / Conducto Arterioso Permeable / Dispositivo Oclusor Septal Tipo de estudio: Diagnostic_studies / Observational_studies Límite: Humans / Infant Idioma: En Revista: Catheter Cardiovasc Interv Asunto de la revista: CARDIOLOGIA Año: 2013 Tipo del documento: Article País de afiliación: Estados Unidos