Poisson noise obscures hypometabolic lesions in PET.
Kerr, Wesley T; Lau, Edward P.
Afiliación
  • Kerr WT; UCLA Department of Biomathematics, Los Angeles, California, USA. wesleytk@ucl.edu
Yale J Biol Med ; 85(4): 541-9, 2012 Dec.
Article en En | MEDLINE | ID: mdl-23239953
ABSTRACT
The technology of fluoro-deoxyglucose positron emission tomography (PET) has drastically increased our ability to visualize the metabolic process of numerous neurological diseases. The relationship between the methodological noise sources inherent to PET technology and the resulting noise in the reconstructed image is complex. In this study, we use Monte Carlo simulations to examine the effect of Poisson noise in the PET signal on the noise in reconstructed space for two pervasive reconstruction algorithms the historical filtered back-projection (FBP) and the more modern expectation maximization (EM). We confirm previous observations that the image reconstructed with the FBP biases all intensity values toward the mean, likely due to spatial spreading of high intensity voxels. However, we demonstrate that in both algorithms the variance from high intensity voxels spreads to low intensity voxels and obliterates their signal to noise ratio. This finding has profound impacts on the clinical interpretation of hypometabolic lesions. Our results suggest that PET is relatively insensitive when it comes to detecting and quantifying changes in hypometabolic tissue. Further, the images reconstructed with EM visually match the original images more closely, but more detailed analysis reveals as much as a 40 percent decrease in the signal to noise ratio for high intensity voxels relative to the FBP. This suggests that even though the apparent spatial resolution of EM outperforms FBP, the signal to noise ratio of the intensity of each voxel may be higher in the FBP. Therefore, EM may be most appropriate for manual visualization of pathology, but FBP should be used when analyzing quantitative markers of the PET signal. This suggestion that different reconstruction algorithms should be used for quantification versus visualization represents a major paradigm shift in the analysis and interpretation of PET images.
Asunto(s)
Palabras clave

Texto completo: 1 Bases de datos: MEDLINE Asunto principal: Ruido Tipo de estudio: Guideline / Health_economic_evaluation Límite: Humans Idioma: En Revista: Yale J Biol Med Año: 2012 Tipo del documento: Article País de afiliación: Estados Unidos

Texto completo: 1 Bases de datos: MEDLINE Asunto principal: Ruido Tipo de estudio: Guideline / Health_economic_evaluation Límite: Humans Idioma: En Revista: Yale J Biol Med Año: 2012 Tipo del documento: Article País de afiliación: Estados Unidos